FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19

MDR report key: 12164306 · Received July 13, 2021

Report

Report Number
1221359-2021-01991
Event Type
Malfunction
Date Received
July 13, 2021
Report Date
August 20, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCABOROUGH
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1023594 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER190-000 / LOT 1023594 AND TEST BASE PART NUMBER 190-430 / LOT 1023594. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1028103 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE AS THE LOGFILES WERE NOT PROVIDED. REVIEW OF COMPLAINTS AGAINST THE KIT LOTS FOR THE REPORTED ISSUE INDICATES PRODUCT PERFORMANCE IS AS EXPECTED AND HAVE NOT IDENTIFIED A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

INVESTIGATION NOT YET COMPLETE. UPON COMPLETION, THE INFORMATION WILL BE PROVIDED IN A SUPPLEMENTAL REPORT. REFERENCE MFR NUMBERS: 1221359-2021-01990, 1221359-2021-01992, 1121359-2021-01963.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY FOR MULTIPLE PATIENTS. THIS MFR. REPORT ADDRESSES PATIENT THREE (3) OF FOUR (4). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY. ADDITIONAL INFORMATION REQUESTED BUT HAS NOT BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1053623 ID NOW COVID-19 MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCABOROUGH 1023594 10811877011269

Patients

Seq Age Sex Outcome Treatment
1