FDA Adverse Event Other Summary report: N

INTERA 1.5T POWER

MDR report key: 1216263 · Received October 23, 2008

Report

Report Number
1217116-2008-00076
Event Type
Other
Date Received
October 23, 2008
Date of Event
August 31, 2008
Report Date
September 15, 2008
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
LNG
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THIS REPORT IS BEING FILED UPON NOTIFICATION FROM OUR X-RAY MFR TO SUBMIT THIS EVENT THAT OCCURRED. PROBLEM: THE PT WAS HAVING AN MRI PROCEDURE. THE PT WAS SCANNED WITH THE SYNERGY BODY COIL (SBC) WITH THE FEET FIRST INTO THE MAGNET. THE SBC WAS PLACED OVER THE ANTERIOR AND POSTERIOR UPPER HIP/THIGH. THE PT COMPLAINED OF A WARMING BETWEEN HIS THIGHS AT THE END OF THE EXAM. ABOUT TWO HOURS AFTER THE EXAMINATION, A SECOND DEGREE BURN WITH A 2 CM BLISTER APPEARED ON ONE LEG. THERE WAS SKIN TO SKIN CONTACT BETWEEN THE TWO LEGS AT THE LOCATION OF THIS BURN. THE PT WAS TREATED WITH BURN OINTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERA 1.5T POWER LNG (MAGNETIC RESONANCE DIAGNOSTIC DEVICE) LNG PHILIPS MEDICAL SYSTEMS 781105 NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR Other