FDA Adverse Event
Other
Summary report: N
INTERA 1.5T POWER
MDR report key: 1216263
·
Received October 23, 2008
Report
- Report Number
- 1217116-2008-00076
- Event Type
- Other
- Date Received
- October 23, 2008
- Date of Event
- August 31, 2008
- Report Date
- September 15, 2008
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- LNG
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THIS REPORT IS BEING FILED UPON NOTIFICATION FROM OUR X-RAY MFR TO SUBMIT THIS EVENT THAT OCCURRED. PROBLEM: THE PT WAS HAVING AN MRI PROCEDURE. THE PT WAS SCANNED WITH THE SYNERGY BODY COIL (SBC) WITH THE FEET FIRST INTO THE MAGNET. THE SBC WAS PLACED OVER THE ANTERIOR AND POSTERIOR UPPER HIP/THIGH. THE PT COMPLAINED OF A WARMING BETWEEN HIS THIGHS AT THE END OF THE EXAM. ABOUT TWO HOURS AFTER THE EXAMINATION, A SECOND DEGREE BURN WITH A 2 CM BLISTER APPEARED ON ONE LEG. THERE WAS SKIN TO SKIN CONTACT BETWEEN THE TWO LEGS AT THE LOCATION OF THIS BURN. THE PT WAS TREATED WITH BURN OINTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INTERA 1.5T POWER | LNG (MAGNETIC RESONANCE DIAGNOSTIC DEVICE) | LNG | PHILIPS MEDICAL SYSTEMS | 781105 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Other |