MUST MINI ROD TI 3.5 X 80MM (STERILE)
Report
- Report Number
- 3005180920-2021-00589
- Event Type
- Injury
- Date Received
- July 13, 2021
- Date of Event
- June 22, 2021
- Report Date
- August 3, 2021
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- NKG
- UDI-DI
- 07630030864988
- PMA / PMN Number
- K171369
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
DEVICES RETURNED ON 21-JULY-2021 AD ANALYZED. VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER. DURING THE VISUAL INSPECTION ON RETRIEVED SET-SCREWS COMPONENT, IT WAS OBSERVED THAT THE COMPONENTS WERE STILL IN GOOD CONDITION WITH NO SIGN OF DEFORMATION AND/OR FAILURE. ADDITIONALLY, A TEST WAS PERFORMED TO INVESTIGATE ABOUT THE POSSIBLE ROOT CAUSE AND VERIFY THE ABILITY OF THE SET-SCREWS TO FULLY SECURE THE OVERALL SCREWS-ROD CONSTRUCT. IN CONCLUSION, WE HYPOSTHESIZE THAT THE ROOT CAUSE OF THIS COMPLAINT IS DUE TO NOT PERFORMING THE FINAL TIGHETENING IN THE PROPER WAY, BY USING THE ASSOCIATED PRE-SET TORQUE LIMITER SCREWDRIVER AS GIVEN IN THE RELATED SURGITAL TECHNIQUE.
BATCH REVIEW PERFORMED ON 23.06.2021: LOT 2120973: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 12-APR-2021. EXPIRATION DATE: 2026-03-04. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT SIMILAR REPORTED EVENT. ADDITIONAL ITEMS INVOLVED IN THE EVENT, BATCH REVIEW PERFORMED ON 23.06.2021: MUST MINI 03.75.600 MUST MINI ROD TI Ø3.5 X 80MM (STERILE) (K171369) LOT. 1922070 LOT 1922070: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 19-07-2019. EXPIRATION DATE: 2024-07-03. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT SIMILAR REPORTED EVENT. MUST MINI 03.75.002 MUST MINI POLYAXIAL SCREW 3.5 X 14 FULL THREAD + NUT (K171369) LOT. 2022678 LOT 2022678: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 22-10-2020. EXPIRATION DATE: 2025-10-07. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH 3 SIMILAR REPORTED EVENTS (ALL IMPLANTS REPORTED IN THIS COMPLAINT). MUST MINI 03.75.002 MUST MINI POLYAXIAL SCREW 3.5 X 14 FULL THREAD + NUT (K171369) LOT. 2022677 LOT 2022677: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 16-11-2020. EXPIRATION DATE: 2025-10-18. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH 1 SIMILAR REPORTED EVENT (REPORTED IN THIS COMPLAINT). MUST MINI 03.75.002 MUST MINI POLYAXIAL SCREW 3.5 X 14 FULL THREAD + NUT (K171369) LOT. 2022255 LOT 2022255: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 07-10-2020. EXPIRATION DATE: 2025-09-20. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT SIMILAR REPORTED EVENT. MUST MINI 03.75.002 MUST MINI POLYAXIAL SCREW 3.5 X 14 FULL THREAD + NUT (K171369) LOT. 2021938 LOT 2021938: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 21-09-2020. EXPIRATION DATE: 2025-09-060. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITHOUT SIMILAR REPORTED EVENT.
C3-C6 FIXATION, SURGERY PERFORMED ON JUNE 16. ON THE NEXT DAY THE SURGEON TOOK THE CT IMAGE AND CONFIRMED THE RODS DISENGAGEMENT FROM THE PEDICLE SCREWS. THE SURGEON CONFIRMED 2 SET SCREWS ON THE C4 COMPLETELY DISENGAGED AND REMAINED 6 SET SCREWS LOOSENED. NO CROSS CONNECTOR USED AT THE PRIMARY SURGERY. ALL IMPLANTS WERE REVISED. THE INSTRUMENTS USED DID NOT SHOW ANY PROBLEM DURING OTHER SURGERIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1060598 | MUST MINI ROD TI 3.5 X 80MM (STERILE) | SPINE ROD | NKG | MEDACTA INTERNATIONAL SA | 03.75.600 | 2120973 | 07630030864988 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Required Intervention |