PEN NDL 32G 4MM
Report
- Report Number
- 9616656-2021-00817
- Event Type
- Malfunction
- Date Received
- July 13, 2021
- Date of Event
- June 10, 2021
- Report Date
- June 21, 2021
- Manufacturer
- BECTON DICKINSON AND CO.
- Product Code
- FMI
- UDI-DI
- 00382903205509
- PMA / PMN Number
- K182320
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. INVESTIGATION CONCLUSION: A COMPLAINT LOT HISTORY CHECK WAS PERFORMED ON LOT # 0253602 FOR NEEDLE CLOG THIS IS THE 3RD. RELATED COMPLAINT FOR NEEDLE CLOG ON LOT # 0253602. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED AS PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND NO CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED AT THIS TIME.
IT WAS REPORTED THAT 2 PEN NDL 32G 4MM WERE UNABLE TO DELIVER INSULIN. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CONSUMER REPORTED FINDING PEN NEEDLES THAT WILL NOT ALLOW LANTUS OUT OF NEEDLES. TOTAL 2 ON THURSDAY (B)(6) 2021.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1057662 | PEN NDL 32G 4MM | PEN NEEDLE | FMI | BECTON DICKINSON AND CO. | 320550 | 0253602 | 00382903205509 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |