FDA Adverse Event Other Summary report: N

TRIUMPH-VTX

MDR report key: 1215641 · Received October 20, 2008

Report

Report Number
1056436-2006-00082
Event Type
Other
Date Received
October 20, 2008
Date of Event
November 6, 2006
Report Date
October 15, 2008
Manufacturer
ANGIODYNAMICS, INC.
Product Code
LJT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS CONFIRMED FOR "THE ATTACHMENT SYSTEM COMING OFF", IT APPEARS THE STEM BECAME DETACHED DUE TO NO GLUE/ADHESIVE. NO MANUFACTURING VARIANCES RELATED TO THIS EVENT OBSERVED IN THE DEVICE HISTORY RECORD REVIEW. A CORRECTIVE ACTION WAS IMPLEMENTED TO CORRECT THIS ISSUE.

Description of Event or Problem · 1

DR. ATTEMPTED TO IMPLANT THIS PORT. THE ATTACHMENT SYSTEM CAME OFF WHILE HE WAS CONNECTING IT TO THE CATHETER. A SECOND PORT WAS OPENED AND USED WITHOUT INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRIUMPH-VTX VASCULAR ACCESS PORT LJT ANGIODYNAMICS, INC. NA 25660

Patients

Seq Age Sex Outcome Treatment
1 UNK Other