Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION. THE CUSTOMER WAS UNABLE TO PROVIDE THE REQUIRED INTAKE INFORMATION. THE INFORMATION TO ENABLE FURTHER INVESTIGATION, SUCH AS THE KIT'S LOT NUMBER, WAS NOT PROVIDED AND AN INVESTIGATION WAS NOT ABLE TO BE PERFORMED. NOTWITHSTANDING, COMPLAINTS AGAINST THESE TREND CODES ARE MONITORED TO IDENTIFY AND TRACK ANY OUT-OF-TREND/UNEXPECTED PERFORMANCE AT THE LOT AND PRODUCT FAMILY LEVEL. IN CONCLUSION, ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUES AS NO INFORMATION WAS PROVIDED FOR INVESTIGATION OR VIGILANCE REPORTING.

Additional Manufacturer Narrative · 1

THE ESTABLISHMENT NAME EXCEEDS THE CHARACTER LIMITS OF INITIAL REPORTER: (B)(6). THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. PLEASE REFERENCE MFR. REPORT #S: 1221359-2021-01939.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 AG CARD USING AN UNSPECIFIED LOT (139666 OR 133584) PERFORMED ON (B)(6) 2021. THIS MFR. REPORT ADDRESSES LOT 2 OF 2. THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE BINAXNOW COVID-19 AG CARD PERFORMED ON (B)(6) 2021 ON A DIRECT TESTED NASAL (ANTERIOR NASAL) SWAB. THE CUSTOMER REPORTED THAT ON (B)(6) 2021 THE PATIENT'S FIRST TESTED NEGATIVE WITH THE BINAXNOW COVID-19 AG CARD. REPEAT TESTING WAS PERFORMED WITH THE BINAXNOW COVID-19 AG CARD ON (B)(6) 2021 USING A NEW SAMPLE AND AGAIN THE PATIENT RESULTS WERE NEGATIVE . PCR CONFIRMATION TESTING (TAQPATH ASSAY) WAS PERFORMED (B)(6) 2021 ON A NASAL SWAB AND GENERATED NEGATIVE RESULTS. THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO TREATMENT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1057130 BINAX NOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 133584

Patients

Seq Age Sex Outcome Treatment
1 Unknown