FDA Adverse Event Malfunction Summary report: N

BD RAPID DETECTION OF SARS-COV-2 VERITOR

MDR report key: 12155706 · Received July 12, 2021

Report

Report Number
3006948883-2021-00765
Event Type
Malfunction
Date Received
July 12, 2021
Date of Event
May 10, 2021
Report Date
September 21, 2021
Manufacturer
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
Product Code
QKP
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE POSITIVE RESULTS WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR (SELECT: MATERIAL # 256082 OR 256089CE), BATCH NUMBER 1012038, 1030316, 1031665. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE POSITIVE COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. AN INVESTIGATION AND TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. PHOTOS WERE PROVIDED AND SHOWED ¿COV 2: +¿ IN THE DISPLAY WINDOW. THE REPORTED ISSUE WAS UNABLE TO BE CONFIRMED. THE ROOT CAUSE COULD NOT BE IDENTIFIED. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS. CAPA 1878253 HAS BEEN INITIATED. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT EUA#: (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "FALSE POSITIVE".

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. EUA#: (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "FALSE POSITIVE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1051976 BD RAPID DETECTION OF SARS-COV-2 VERITOR CORONAVIRUS ANTIGEN DETECTION SYSTEM QKP BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. 1031665

Patients

Seq Age Sex Outcome Treatment
1