FDA Adverse Event Death Summary report: N

APTIO AUTOMATION

MDR report key: 12150638 · Received July 12, 2021

Report

Report Number
3010825766-2021-00008
Event Type
Death
Date Received
July 12, 2021
Date of Event
May 11, 2021
Report Date
September 13, 2021
Manufacturer
INPECO SA
Product Code
DAP
UDI-DI
07640172342008
PMA / PMN Number
K121012
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INITIAL REPORT (3010825766-2021-00008) HAS BEEN SUBMITTED ON JULY 12, 2021. A SUPPLEMENTAL REPORT (3010825766-2021-00008_S1) HAS BEEN SUBMITTED ON AUGUST 2, 2021. SUPPLEMENTAL REPORT: NO ADDITIONAL FINDINGS REGARDING THE INVESTIGATION HAVE BEEN COMMUNICATED BY THE DISTRIBUTOR. THE INVESTIGATION HAS BEEN COMPLETED WITH THE CONFIRMATION THAT THE AUTOMATION SYSTEM DESIGN IS ADEQUATE. NO FURTHER ACTIONS ARE FORESEEN.

Additional Manufacturer Narrative · 1

THE INITIAL REPORT (3010825766-2021-00008) HAS BEEN SUBMITTED ON JULY 12, 2021. SUPPLEMENTAL REPORT: PER FDA REQUEST, THE PRODUCT CODE HAS BEEN CORRECTED TO DAP. ("FIBRINOGEN/FIBRIN DEGRADATION PRODUCTS ASSAY") WHICH IS THE TEST PRODUCT CODE FOR WHICH THE INACCURATE RESULTS WERE GENERATED. ADDITIONAL INFORMATION RECEIVED FROM THE DISTRIBUTOR ABOUT THE DECEASED PATIENTS. THE INITIAL REPORTER SUBMITTED THE MDR 3007494875-2021-00001 ON 21-JUL-2021. NO ADDITIONAL FINDINGS REGARDING THE INVESTIGATION HAVE BEEN COMMUNICATED BY THE DISTRIBUTOR.

Additional Manufacturer Narrative · 1

ACCORDING TO THE ANALYSIS PROVIDED BY THE DISTRIBUTOR, THE MISSED TRANSMISSION OF THE FLAGS ASSOCIATED TO THE TEST RESULTS WAS CAUSED BY THE WRONG CONFIGURATION OF THE STAGO STA R MAX INTERFACE MODULE BY THE DISTRIBUTOR SERVICE PERSONNEL WHO INSTALLED THE WRONG DRIVER, THE DRIVER REQUIRED TO INTERFACE DMS WITH STAGO STA R INSTEAD OF THE DRIVER REQUIRED FOR STAGO STA R MAX. THE CONFIGURATION ERROR COULD HAVE BEEN DETECTED BY THE SERVICE PERSONNEL DURING THE TESTS AFTER THE INSTALLATION OR BY THE USERS SINCE THE ANALYZER NAME IS DISPLAYED ON THE GRAPHIC USER INTERFACE. THE DOCUMENTATION MADE AVAILABLE TO THE SERVICE PERSONNEL FOR THE INTERFACE MODULE CONFIGURATION AND DRIVER INSTALLATION HAS BEEN REVIEWED AND CONFIRMED TO BE CLEAR AND COMPLETE. THE D-DIMER TEST RESULT IS USUALLY INTERPRETED ALONG WITH OTHER TEST RESULTS TO OBTAIN THE OVERALL CLINICAL PICTURE. THE D-DIMER TEST RESULT BY ITSELF SHOULD NOT LEAD TO ANY THERAPEUTIC DECISION. A TEST RESULT OF 0 IS NOT BELIEVABLE AND SHOULD TRIGGER A RERUN IN THE VALIDATION PHASE. ACCORDING TO THE EVALUATION OF THE HAEMATOLOGIST ON SITE THE PATIENTS DEATH WAS NOT CAUSED BY THE AUTOMATION SYSTEM ERRONEOUS RESULTS TRANSMISSION. ACCORDING TO THE CURRENT INFORMATION, THE AUTOMATION SYSTEM HAS BEEN CONFIRMED NOT TO BE THE CAUSE OF THE INCIDENT. THE INVESTIGATION HAS BEEN DONE BASED ON THE INFORMATION PROVIDED BY THE DISTRIBUTOR. THE DISTRIBUTOR HAS NOT PROVIDED YET INPECO WITH THE LOG FILES NECESSARY TO CONFIRM THEIR ANALYSIS.

Description of Event or Problem · 1

THE AUTOMATION SYSTEM DATA MANAGEMENT SOFTWARE (DMS) HAS SENT TO THE UPPER MIDDLEWARE INCORRECT D-DIMER TEST RESULTS ASSOCIATED TO 13 SAMPLE TUBES. THE ANALYZER STA R MAX, MANUFACTURED BY STAGO, SENT TO DMS THE D-DIMER TEST RESULTS > 20 WITH THE FLAG M (MAX), DMS DID NOT TRANSMIT THE FLAG AND REPORTED TO THE UPPER MIDDLEWARE THE OUT OF RANGE RESULTS AS 0. ACCORDING TO THE INFORMATION RECEIVED FROM THE FIELD 3 OF THESE PATIENTS PASSED AWAY. HOWEVER THE LABORATORY HAEMATOLOGIST SPECIFIED THAT THE D-DIMER RESULTS DID NOT CAUSE THE PATIENTS' DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1047642 APTIO AUTOMATION LABORATORY AUTOMATION SYSTEM DAP INPECO SA AP2 07640172342008

Patients

Seq Age Sex Outcome Treatment
1 Death