FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 12148028 · Received July 9, 2021

Report

Report Number
1645337-2021-07751
Event Type
Injury
Date Received
July 9, 2021
Date of Event
June 14, 2021
Report Date
June 16, 2021
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001294
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 2-AUG-2021, MENTOR RECEIVED ADDITIONAL INFORMATION REGARDING THE SUSPECTED MEDICAL DEVICE AND IMPLANTATION DATE. INITIALLY, THE SUSPECTED MEDICAL DEVICE WAS REPORTED AS A 475CC MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS. HOWEVER, ADDITIONAL INFORMATION REVEALED THAT THE ACTUAL DEVICE IS A 575CC MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESES (CATALOG #: 3501690, LOT #: 5990660). A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. INITIALLY, THE DATE OF IMPLANTATION WAS REPORTED AS (B)(6) 2011. HOWEVER, ADDITIONAL INFORMATION REVEALED THAT THE ACTUAL DATE OF IMPLANTATION WAS (B)(6) 2013. ON 6-AUG-2021, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE RELEVANT FIELDS HAVE BEEN UPDATED. MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON AUGUST 21, 2021, IT WAS FOUND THAT ADDITIONAL INFORMATION REGARDING THE PROCEDURE TYPE WAS INADVERTENTLY OMITTED FROM THE PREVIOUS REPORT. ON AUGUST 25, 2021, DEVICE EVALUATION WAS COMPLETED. DEVICE EVALUATION SUMMARY: MENTOR CONDUCTED A VISUAL INSPECTION, LEAK TESTING, MICROSCOPIC EXAMINATION, AND THICKNESS MEASUREMENT OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THE SAL SMOOTH RND DIAP 575CC BREAST IMPLANT WAS FOUND TO HAVE A TEAR ON THE ANTERIOR VIEW, NEAR THE VALVE SYSTEM. LEAK TESTING WAS PERFORMED, IN ACCORDANCE WITH MENTOR PROCEDURES, AND NO MORE LEAK SITES WERE DETECTED DURING THE ANALYSIS. MICROSCOPIC EXAMINATION WAS PERFORMED AND THE CAUSE OF THE DEFLATION COULD NOT BE IDENTIFIED. THEREFORE A THICKNESS MEASUREMENT WAS CONDUCTED ON THE SHELL AT THE TEAR, AND THE THICKNESS WAS WITHIN MANUFACTURING SPECIFICATIONS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: ACCORDING TO THE PRODUCT INSERT DATA SHEET, THE VALVE SYSTEM CAN BE DAMAGED BY IMPROPER USE OF THE FILL-TUBE STYLET. CARE SHOULD BE TAKEN THAT THE STYLET ENTERS THE VALVE SMOOTHLY. USE THE THUMB AND FOREFINGER TO STABILIZE/SUPPORT THE VALVE SEAT AND GENTLY PUSH THE STYLET TIP INTO THE VALVE OPENING. OVERSTRESSING THE VALVE MATERIAL MAY RESULT IN PUNCTURES OR TEARS AND SUBSEQUENT DEFLATION MAY OCCUR. USE ONLY THE FILL TUBE STYLET PROVIDED WITH THIS PRODUCT. TAKE CARE NOT TO PUNCTURE THE DIAPHRAGM VALVE OR THE SHELL WITH THE STYLET TIP. CARE MUST ALSO BE TAKEN WHEN THE FILL TUBE STYLET IS REMOVED TO PREVENT DAMAGE TO THE VALVE ASSEMBLY. AS PART OF MENTOR¿S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. MANUFACTURER'S REFERENCE NUMBER: (B)(4) INITIALLY, IT WAS REPORTED THAT THE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH THE SUSPECTED MEDICAL DEVICE. HOWEVER, THE PATIENT UNDERWENT BREAST AUGMENTATION WITH THE SUSPECTED COMPLAINT DEVICE.

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: DEFLATION. DATE OF EXPLANTATION: (B)(6) 2021. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) CAUCASIAN FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH A 475CC MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS AND EXPERIENCED POSTOPERATIVE RIGHT-SIDED DEFLATION. THE DIAGNOSIS WAS CONFIRMED BY A HEALTHCARE PROFESSIONAL. AS A RESULT, THE PATIENT IS SCHEDULED TO UNDERGO EXPLANTATION ON (B)(6) 2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1041913 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 3501680 5990660 00081317001294

Patients

Seq Age Sex Outcome Treatment
1 36 YR Required Intervention