FDA Adverse Event Malfunction Summary report: N

BD FLU+ SYRINGE

MDR report key: 12147421 · Received July 9, 2021

Report

Report Number
3002682307-2021-00314
Event Type
Malfunction
Date Received
July 9, 2021
Date of Event
June 1, 2021
Report Date
August 23, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
K091377
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2012414. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR THIS INCIDENT, TWENTY RETAINED SAMPLES OF THE SAME LOT NUMBER WERE OBTAINED FROM THE MANUFACTURING FACILITY FOR FURTHER EVALUATION. THE RETAINED SAMPLES WERE EXAMINED AND NO SIGNS OF LEAKAGE WERE IDENTIFIED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE RELATED TO THE MANUFACTURING PROCESS COULD NOT BE DETERMINED FOR THIS INCIDENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT VACCINE MEDICINE LEAKED FROM THE BOTTOM OF THE BD FLU+¿ SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "VACCINE NOTED TO BE LEAKING OUT OF THE BOTTOM OF THE SYRINGE, NO INJURY, SYRINGE FILLED WITH VACCINE DISPOSED OF."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT VACCINE MEDICINE LEAKED FROM THE BOTTOM OF THE BD FLU+¿ SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "VACCINE NOTED TO BE LEAKING OUT OF THE BOTTOM OF THE SYRINGE, NO INJURY, SYRINGE FILLED WITH VACCINE DISPOSED OF."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1040640 BD FLU+ SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2012414

Patients

Seq Age Sex Outcome Treatment
1