FDA Adverse Event
Malfunction
Summary report: N
IMPACTION BUR GUARD
MDR report key: 1214046
·
Received October 23, 2008
Report
- Report Number
- 1811755-2008-00077
- Event Type
- Malfunction
- Date Received
- October 23, 2008
- Date of Event
- September 26, 2008
- Report Date
- September 26, 2008
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- EJL
- PMA / PMN Number
- K961970
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE HAND PIECE AND BUR GUARD WERE RECEIVED AT THE MANUFACTURER FOR TESTING, AND THE ALLEGED CONDITION OF THE BUR GUARD MELTING ONTO THE HAND PIECE WAS CONFIRMED. THE HAND PIECE PASSED THE QUALITY INSPECTION PROCEDURE ACCORDING TO SPECIFICATION. THE BUR GUARD CAN MELT IF IT IS PUSHED UP ONTO THE HAND PIECE. WHEN THIS HAPPENS, THE NOSE CONE COMES INTO CONTACT WITH THE BUR GUARD, AND THE FRICTION CAN MELT THE BUR GUARD.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN ORAL PROCEDURE, THE BUR GUARD MELTED ONTO THE NOSE CONE OF THE HANDPIECE. THERE WAS NO PT OR USER INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMPACTION BUR GUARD | BONE CUTTING INSTRUMENT AND ACCESSORIES | EJL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |