FDA Adverse Event Malfunction Summary report: N

IMPACTION BUR GUARD

MDR report key: 1214046 · Received October 23, 2008

Report

Report Number
1811755-2008-00077
Event Type
Malfunction
Date Received
October 23, 2008
Date of Event
September 26, 2008
Report Date
September 26, 2008
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
EJL
PMA / PMN Number
K961970
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE HAND PIECE AND BUR GUARD WERE RECEIVED AT THE MANUFACTURER FOR TESTING, AND THE ALLEGED CONDITION OF THE BUR GUARD MELTING ONTO THE HAND PIECE WAS CONFIRMED. THE HAND PIECE PASSED THE QUALITY INSPECTION PROCEDURE ACCORDING TO SPECIFICATION. THE BUR GUARD CAN MELT IF IT IS PUSHED UP ONTO THE HAND PIECE. WHEN THIS HAPPENS, THE NOSE CONE COMES INTO CONTACT WITH THE BUR GUARD, AND THE FRICTION CAN MELT THE BUR GUARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ORAL PROCEDURE, THE BUR GUARD MELTED ONTO THE NOSE CONE OF THE HANDPIECE. THERE WAS NO PT OR USER INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMPACTION BUR GUARD BONE CUTTING INSTRUMENT AND ACCESSORIES EJL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK