FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722NAB PRDGM INS V2.2 BL EN

MDR report key: 1214043 · Received October 23, 2008

Report

Report Number
3004209178-2008-00886
Event Type
Malfunction
Date Received
October 23, 2008
Date of Event
October 10, 2008
Report Date
October 10, 2008
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
LZG
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DUE TO THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE. HOWEVER, THE INSULIN PUMP PASSED THE BASIC OCCLUSION, PRIME, AND EXCESSIVE NO DELIVERY ALARM TESTS.

Description of Event or Problem · 1

THE CUSTOMER COMPLAINED OF THE INSULIN PUMP ALARMING MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAB PRDGM INS V2.2 BL EN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAB

Patients

Seq Age Sex Outcome Treatment
1 43 YR