FDA Adverse Event Malfunction Summary report: N

PUMP MMT-722RNAB PRDGM INS BL EN RC

MDR report key: 1214023 · Received October 23, 2008

Report

Report Number
2032227-2008-01847
Event Type
Malfunction
Date Received
October 23, 2008
Date of Event
October 15, 2008
Report Date
October 15, 2008
Manufacturer
MEDTRONIC MINIMED
Product Code
LZG
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR, DUE TO THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. THE CUSTOMER WAS ADVISED TO REVERT TO A BACKUP PLAN TO TREAT HER DIABETES. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722RNAB PRDGM INS BL EN RC INFUSION PUMP LZG MEDTRONIC MINIMED MMT-722RNAB

Patients

Seq Age Sex Outcome Treatment
1 71 YR