FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-722RNAB PRDGM INS BL EN RC
MDR report key: 1214023
·
Received October 23, 2008
Report
- Report Number
- 2032227-2008-01847
- Event Type
- Malfunction
- Date Received
- October 23, 2008
- Date of Event
- October 15, 2008
- Report Date
- October 15, 2008
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- LZG
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR, DUE TO THE MOTOR ENCODER SIGNAL BEING OUT OF PHASE.
Description of Event or Problem · 1
THE CUSTOMER STATED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. THE CUSTOMER WAS ADVISED TO REVERT TO A BACKUP PLAN TO TREAT HER DIABETES. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-722RNAB PRDGM INS BL EN RC | INFUSION PUMP | LZG | MEDTRONIC MINIMED | MMT-722RNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |