FDA Adverse Event Malfunction Summary report: N

DIMENSION CLINCAL CHEMISTRY SYSTEM

MDR report key: 1213984 · Received October 24, 2008

Report

Report Number
1226181-2008-00113
Event Type
Malfunction
Date Received
October 24, 2008
Date of Event
September 26, 2008
Report Date
September 29, 2008
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC
Product Code
MMI
PMA / PMN Number
K944093
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS HEALTHCARE DIAGNOSTICS, INC FIELD SERVICE REPRESENTATIVE (FSE) WAS DISPATCHED TO THE ACCOUNT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS REAGENT WICKING TO ADJACENT CUVETTES. THE FSE CORRECTED THE MALFUNCTION. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF FALSELY ELEVATED TROPONIN I RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION CLINCAL CHEMISTRY SYSTEM CLINICAL CHEMISTRY SYSTEM MMI SIEMENS HEALTHCARE DIAGNOSTICS, INC RXLMAX NA

Patients

Seq Age Sex Outcome Treatment
1