FDA Adverse Event
Malfunction
Summary report: N
ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER
MDR report key: 1213689
·
Received October 23, 2008
Report
- Report Number
- 2084725-2008-00704
- Event Type
- Malfunction
- Date Received
- October 23, 2008
- Date of Event
- September 26, 2008
- Report Date
- September 26, 2008
- Manufacturer
- MINNTECH CORPORATION
- Product Code
- NVE
- PMA / PMN Number
- K832458
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
OTHER - CAPITAL EQUIPMENT, EVALUATED AT CUSTOMER SITE. THE FSE FOUND HEATER FUSE ON MICROPROCESSOR BOARD BAD. THE FSE REPLACED FUSE AND WAITED FOR TEMPERATURE LIGHT TO COME ON. TEMP LIGHT CAME ON WITHIN 10 MINUTES.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE TEMPERATURE LIGHT WAS NOT ON. ASP CUSTOMER CARE ADVISED THE CUSTOMER TO CHECK THE TEMPERATURE WITH AN EXTERNAL THERMOMETER. THERE WERE NO PHYSICAL COMPLAINTS REPORTED. THE ASP FIELD SERVICE ENGINEER (FSE) WENT TO THE FACILITY TO ASSESS THE UNIT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER | ENDOSCOPE REPROCESSOR | NVE | MINNTECH CORPORATION | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |