FDA Adverse Event Malfunction Summary report: N

ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER

MDR report key: 1213689 · Received October 23, 2008

Report

Report Number
2084725-2008-00704
Event Type
Malfunction
Date Received
October 23, 2008
Date of Event
September 26, 2008
Report Date
September 26, 2008
Manufacturer
MINNTECH CORPORATION
Product Code
NVE
PMA / PMN Number
K832458
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OTHER - CAPITAL EQUIPMENT, EVALUATED AT CUSTOMER SITE. THE FSE FOUND HEATER FUSE ON MICROPROCESSOR BOARD BAD. THE FSE REPLACED FUSE AND WAITED FOR TEMPERATURE LIGHT TO COME ON. TEMP LIGHT CAME ON WITHIN 10 MINUTES.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE TEMPERATURE LIGHT WAS NOT ON. ASP CUSTOMER CARE ADVISED THE CUSTOMER TO CHECK THE TEMPERATURE WITH AN EXTERNAL THERMOMETER. THERE WERE NO PHYSICAL COMPLAINTS REPORTED. THE ASP FIELD SERVICE ENGINEER (FSE) WENT TO THE FACILITY TO ASSESS THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASP AUTOMATIC ENDOSCOPE REPROCESSOR WITH PRINTER ENDOSCOPE REPROCESSOR NVE MINNTECH CORPORATION NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA