FDA Adverse Event Other Summary report: N

DIMENSION CLINICAL CHEMISTRY SYSTEM

MDR report key: 1213684 · Received October 24, 2008

Report

Report Number
2517506-2008-00096
Event Type
Other
Date Received
October 24, 2008
Date of Event
October 2, 2008
Report Date
October 3, 2008
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Product Code
MMI
PMA / PMN Number
K010313
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS SAMPLE INTEGRITY. THE IFU FOR DIMENSION CTNI FLEX REAGENT CARTRIDGE CONTAINS THE FOLLOWING INFORMATION: "SPECIMENS SHOULD BE FREE OF PARTICULATE MATTER TO PREVENT APPEARANCE OF FIBRIN IN SERUM SAMPLES, COMPLETE CLOT FORMATION SHOULD TAKE PLACE BEFORE CENTRIFUGATION." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. PT UNDERWENT A CARDIAC CATHETERIZATION. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION CLINICAL CHEMISTRY SYSTEM CTNI FLEX REAGENT CARTRIDGE MMI SIEMENS HEALTHCARE DIAGNOSTICS, INC. NA DB9204

Patients

Seq Age Sex Outcome Treatment
1 78 YR