FDA Adverse Event Malfunction Summary report: N

MAXPLUS POSITIVE PRESSURE CONNECTOR

MDR report key: 12132323 · Received July 7, 2021

Report

Report Number
9616066-2021-51467
Event Type
Malfunction
Date Received
July 7, 2021
Date of Event
June 7, 2021
Report Date
June 22, 2021
Manufacturer
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
Product Code
FPA
PMA / PMN Number
K072542
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MP1000 CHINA 510K: THIS IS AN INTERNATIONAL CODE. THE MODEL#/CATALOG# IDENTIFIED IS A CAREFUSION PRODUCT WHICH IS SAME OR SIMILAR TO A DEVICE THAT IS APPROVED FOR SALE DOMESTICALLY. THE DOMESTIC SIMILAR LIST NUMBER IS MP1000. THE 510K NUMBER PROVIDED IS FOR THE DOMESTIC SIMILAR PRODUCT: K072542. INVESTIGATION SUMMARY: AN MP1000 CHINA PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION; HOWEVER THE CUSTOMER CONFIRMED THAT THE COMPLAINT SAMPLE WAS FROM LOT 20076476. ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER CONFIRMED THE OCCLUSION WAS IDENTIFIED WHILST CONNECTED TO A RONGDE LUER SLIP SYRINGE DURING THE PROCESS OF AN AIR EXHAUST. THE DETAILS OF THIS FEEDBACK WERE FORWARDED TO THE MANUFACTURING SITE FOR INVESTIGATION. A REVIEW OF THE PRODUCTION RECORDS FOR LOT 20076476 DID NOT IDENTIFY ANY IN-PROCESS TESTING FAILURES OR QUALITY DEVIATIONS WHICH MAY HAVE RESULTED IN A REPORT OF THIS NATURE. IT WAS NOT POSSIBLE TO CONFIRM THE ROOT CAUSE OF THE REPORTED ISSUE IN THIS INSTANCE, WITHOUT A SAMPLE IT IS NOT POSSIBLE TO DETERMINE WHETHER A MANUFACTURING DEFECT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED OCCLUSION. A REVIEW OF THE CUSTOMER FEEDBACK DATABASE INDICATES THAT COMPLAINTS OF THIS NATURE ARE RARE AND THERE IS CURRENTLY NO TREND FOR ISSUES OF THIS NATURE AGAINST THE MP1000 CHINA PRODUCT.

Description of Event or Problem · 1

IT WAS REPORTED THAT 2 MAXPLUS POSITIVE PRESSURE CONNECTORS WERE BLOCKED DURING NEUROSURGERY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "NEUROSURGERY, NEEDLE FREE CONNECTOR TUBING WERE BLOCKED."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1023905 MAXPLUS POSITIVE PRESSURE CONNECTOR INTRAVASCULAR ADMINISTRATION SET FPA SISTEMAS MEDICOS ALARIS, S.A. DE C.V. 20076476

Patients

Seq Age Sex Outcome Treatment
1