FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 1213118 · Received October 23, 2008

Report

Report Number
1644487-2008-02547
Event Type
Malfunction
Date Received
October 23, 2008
Date of Event
August 20, 2008
Report Date
September 29, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLASHCARD FILE WAS UNABLE TO BE UPLOADED DUE TO AN SQL ERROR. THE FLASHCARD WAS SENT TO THE PRODUCT ANALYSIS LAB FOR EVALUATION. UPON COMPLETION OF PRODUCT TESTING, IT WAS DETERMINED THAT THE EVENT WAS CAUSED BY A CORRUPT DATABASE. THE ANALYSIS COULD NOT DETERMINE THE INITIAL CAUSE OF THE DATABASE CORRUPTION, BUT THAT IT WAS MOST LIKELY ASSOCIATED WITH THE FLASHCARD. NO FURTHER ANOMALIES ASSOCIATED WITH THE FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE UNKNOWN LYJ CYBERONICS, INC. MODEL 250 366994

Patients

Seq Age Sex Outcome Treatment
1