FG 6301 DUAL CHNL VOL INF PUMP
Report
- Report Number
- 6000001-2007-81052
- Event Type
- Malfunction
- Date Received
- October 29, 2008
- Date of Event
- July 3, 2007
- Report Date
- July 5, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- PMA / PMN Number
- K915523
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM THE FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA JULY 31, 2007. THE FACILITY REPRESENTATIVE REPORTED A ¿RIGHT DOOR BROKEN¿ DURING BIO-MED TESTING. EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. THE REPORTED CONDITION OF ¿RIGHT DOOR BROKEN" WAS CONFIRMED. PUMP #1 DOOR ASSEMBLY WAS FOUND CRACKED. THE DOOR ASSEMBLY WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, PUMP #1 DOOR ASSEMBLY WAS FOUND CRACKED. ACCORDING TO THE FACILITY REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FG 6301 DUAL CHNL VOL INF PUMP | 80FRN | FRN | BAXTER HEALTHCARE (SG) | N/A | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |