FDA Adverse Event
Malfunction
Summary report: N
CONTOUR TEST STRIPS (50)
MDR report key: 1212755
·
Received October 20, 2008
Report
- Report Number
- 1826988-2008-01206
- Event Type
- Malfunction
- Date Received
- October 20, 2008
- Date of Event
- October 3, 2008
- Report Date
- October 3, 2008
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE ADVOCATE STATED THAT SHE TESTED THE CUSTOMER'S BLOOD GLUCOSE AND RECEIVED A READING OF "HI" USING THE CUSTOMER'S METER. SHE RETESTED THE CUSTOMER USING HER OWN CONTOUR METER AND RECEIVED A READING OF 171 MG/DL. THE CUSTOMER'S METER READING WAS OFF THE CHART, BUT THE DIFFERENCE BETWEEN THE READINGS APPEARS TO FALL IN THE "C" ZONE OF THE PARKES ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. THE ADVOCATE ALSO PERFORMED CONTROL TESTS AND RECEIVED A RESULT OF 267 MG/DL. THE NORMAL CONTROL RANGE WAS 104-144 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR TEST STRIPS (50) | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE, LLC | 7097B | 7EC3D01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |