FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1212755 · Received October 20, 2008

Report

Report Number
1826988-2008-01206
Event Type
Malfunction
Date Received
October 20, 2008
Date of Event
October 3, 2008
Report Date
October 3, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE ADVOCATE STATED THAT SHE TESTED THE CUSTOMER'S BLOOD GLUCOSE AND RECEIVED A READING OF "HI" USING THE CUSTOMER'S METER. SHE RETESTED THE CUSTOMER USING HER OWN CONTOUR METER AND RECEIVED A READING OF 171 MG/DL. THE CUSTOMER'S METER READING WAS OFF THE CHART, BUT THE DIFFERENCE BETWEEN THE READINGS APPEARS TO FALL IN THE "C" ZONE OF THE PARKES ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. THE ADVOCATE ALSO PERFORMED CONTROL TESTS AND RECEIVED A RESULT OF 267 MG/DL. THE NORMAL CONTROL RANGE WAS 104-144 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC 7097B 7EC3D01

Patients

Seq Age Sex Outcome Treatment
1 UNK