FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1212685 · Received October 20, 2008

Report

Report Number
1826988-2008-01229
Event Type
Malfunction
Date Received
October 20, 2008
Date of Event
October 10, 2008
Report Date
October 10, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

THE CUSTOMER STATED THAT SHE RECEIVED SOME CONTROL TEST RESULTS THAT WERE LOW, OUT OF SPECIFICATION. WHILE TROUBLESHOOTING, SHE PERFORMED 2 CONTROL TESTS AND RECEIVED RESULTS OF 50 AND 47 MG/DL. THE NORMAL CONTROL RANGE WAS 99-136 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC 7080D 7HC3C12

Patients

Seq Age Sex Outcome Treatment
1 UNK