FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 29GA 1/2IN

MDR report key: 12125787 · Received July 6, 2021

Report

Report Number
1920898-2021-00724
Event Type
Malfunction
Date Received
July 6, 2021
Date of Event
June 8, 2021
Report Date
July 12, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K024112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: CUSTOMER RETURNED TWO IMAGES OF THREE 0.3ML SYRINGES. NO OTHER IDENTIFICATION WAS AVAILABLE. THE SYRINGES HAVE HAD THEIR NEEDLE SHIELDS AND HUBS SEPARATE FROM THE BARRELS. THE HUBS HAVE BECOME LODGED INSIDE THE SHIELDS. THERE IS NO DAMAGE TO EITHER THE CONNECTORS AT THE DISTAL TIPS OF THE BARRELS OR THEIR RESPECTIVE NEEDLE HUBS. NO SIGNS OF USE AND NO OTHER DEFECTS FOUND. DUE TO THE BATCH BEING UNKNOWN, NO DHR REVIEW CAN BE COMPLETED. BASED ON THE IMAGES RECEIVED, BD WAS ABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE OF NEEDLE HUB SEPARATION. CAPA#1630423 HAS BEEN OPENED TO ADDRESS THIS ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT 1 BD SYRINGE 0.3ML 29GA 1/2IN HUB SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CUSTOMER REPORTED THAT WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE EXPIRATION DATE : UNKNOWN. INITIAL REPORTER PHONE# : UNKNOWN. INITIAL REPORTER ZIP CODE : UNKNOWN. A DEVICE EVALUATION IS PENDING BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. DEVICE MANUFACTURE DATE : UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT 1 BD SYRINGE 0.3ML 29GA 1/2IN HUB SEPARATED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER : THE CUSTOMER REPORTED THAT WHEN REMOVING THE SHIELD, THE NEEDLE WITH THE SHIED WAS SEPARATED FROM THE BARREL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1021254 SYRINGE 0.3ML 29GA 1/2IN PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1