INFANT CPAP NASAL PRONG
Report
- Report Number
- 9611451-2008-00620
- Event Type
- Injury
- Date Received
- October 24, 2008
- Date of Event
- July 15, 2008
- Report Date
- September 24, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD.
- Product Code
- BZO
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE INVESTIGATION WAS CONDUCTED BASED ON THE EVENT DESCRIPTION AND RESULTS OF PREVIOUS INVESTIGATIONS, AS THE COMPLAINT DEVICE HAD NOT BEEN RETURNED FOR INVESTIGATION. RESULTS: BASED ON THE INFORMATION PROVIDED BY THE HOSPITAL, THE NASAL PRONGS WERE NOT INSERTED ACCORDING TO FISHER & PAYKEL HEALTHCARE'S USER INSTRUCTIONS. IN ADDITION, THE HOSPITAL FURTHER REPORTED THAT THE INFANT'S SKIN INTEGRITY IN THE SEPTAL AREA WAS NOT CHECKED WITH ENOUGH FREQUENCY. CONCLUSION: IT IS WELL KNOWN (THROUGH PUBLISHED LITERATURE), THAT NASAL CPAP REQUIRES CAREFUL MANAGEMENT TO AVOID THE POTENTIAL SIDE EFFECT OF SEPTAL EROSION, AS THE TISSUE AROUND THE NASAL SEPTUM CAN BREAK DOWN IF SUBJECTED TO CONTINUOUS PRESSURE, FRICTION OR MOISTURE. OUR USER INSTRUCTIONS SUPPLIED WITH THE CPAP INTERFACE ILLUSTRATE THE CORRECT INTERFACE SET-UP ON THE PATIENT AND CLEARLY DIRECT USERS TO ENSURE THAT THE NASAL PRONGS ARE POSITIONED WITH AT LEAST A 2 MM GAP FROM THE SEPTUM TO AVOID PRESSURE NECROSIS. THE NASAL SEPTAL DAMAGE CAN BE PREVENTED WITH APPROPRIATE CARE AND ATTENTION FROM MEDICAL PROFESSIONALS. OUR MONITORING AND TRENDING FOR INFANT CPAP NASAL SEPTAL DAMAGE HAS A RATE OF OCCURRENCE WORLDWIDE FOR THE LAST YEAR OF 37 PPM.
A HOSPITAL REPORTED THE FOLLOWING: AN INFANT IN THEIR NICU WAS RECEIVING NASAL CPAP THERAPY VIA AN AVILLA VENTILATOR USING THE FISHER & PAYKEL HEALTHCARE INTERFACE WITH AN INFANT CPAP NASAL PRONG. OVER THE COURSE OF 12 HOURS AFTER PLACING THE INFANT ON THAT SET-UP, THE SEPTUM DEVELOPED A PRESSURE SORE WHICH TURNED NECROTIC AND THE INFANT LOST A PORTION OF THE NASAL SEPTUM; AND THE NASAL PRONGS WERE NOT INSERTED ACCORDING TO THE MANUFACTURER'S USER INSTRUCTIONS AND THE INFANT'S SKIN INTEGRITY IN THE SEPTAL AREA WAS NOT ASSESSED WITH ENOUGH FREQUENCY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT CPAP NASAL PRONG | BZO | FISHER & PAYKEL HEALTHCARE LTD. | BC4030 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 6 MO | Disability | AVILLA VENTILATOR |