FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 12125287 · Received July 6, 2021

Report

Report Number
9617229-2021-46803
Event Type
Injury
Date Received
July 6, 2021
Date of Event
January 20, 2021
Report Date
September 23, 2021
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN ASSISTANT

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS RECEIVED ON AUGUST 12, 2021 WITH LOT NUMBER 2900844. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED WEAR ABRASION, VOIDS AND YELLOW ENCAPSULATION. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE WAS WITHIN SPECIFICATION. CLOUDY AFTER AUTOCLAVE DISINFECTION PROCESS IN THE GEL. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO ISSUES FOUND RELATED WITH THE MANUFACTURING.

Description of Event or Problem · 0

PHYSICIAN REPORTED GRADE IV CAPSULAR CONTRACTURE. LEFT SIDE. DEVICE EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. PHOTO ANALYSIS EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED: WHITE TISSUE, WHITE PARTICLE MATERIAL ON THE SHELL. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE IV.

Description of Event or Problem · 1

PHYSICIAN REPORTED GRADE IV CAPSULAR CONTRACTURE. LEFT SIDE. DEVICE EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1022835 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2900844

Patients

Seq Age Sex Outcome Treatment
1 29 YR Required Intervention