INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2021-46803
- Event Type
- Injury
- Date Received
- July 6, 2021
- Date of Event
- January 20, 2021
- Report Date
- September 23, 2021
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN ASSISTANT
Narratives
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE WAS RECEIVED ON AUGUST 12, 2021 WITH LOT NUMBER 2900844. VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED WEAR ABRASION, VOIDS AND YELLOW ENCAPSULATION. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE WAS WITHIN SPECIFICATION. CLOUDY AFTER AUTOCLAVE DISINFECTION PROCESS IN THE GEL. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: NO ISSUES FOUND RELATED WITH THE MANUFACTURING.
PHYSICIAN REPORTED GRADE IV CAPSULAR CONTRACTURE. LEFT SIDE. DEVICE EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. PHOTO ANALYSIS EVALUATION: VISUAL ANALYSIS OF THE PHOTOGRAPHS IDENTIFIED: WHITE TISSUE, WHITE PARTICLE MATERIAL ON THE SHELL. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS IT IS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE. THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE IV.
PHYSICIAN REPORTED GRADE IV CAPSULAR CONTRACTURE. LEFT SIDE. DEVICE EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1022835 | INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2900844 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR | Required Intervention |