FDA Adverse Event Malfunction Summary report: N

TSRH 3DX

MDR report key: 12119671 · Received July 6, 2021

Report

Report Number
1030489-2021-00891
Event Type
Malfunction
Date Received
July 6, 2021
Date of Event
June 17, 2021
Report Date
September 27, 2021
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
HXB
UDI-DI
00613994282668
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D9, H3, H6 H3. DEVICE EVALUATION SUMMARY: PRODUCT ANALYSIS #704455607:8350294 LOT# KH18J868 VISUAL OBSERVATION REVEALS THAT THE TIP OF THE PROBE HAS BEEN BENT FROM WHAT APPEARS TO BE OVERLOAD. THIS IS CONSISTENT WITH BEND STRESS OVERLOAD. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PMA/510(K)#: CLASS I EXEMPT. NEITHER THE DEVICE NOR FILMS OF APPLICABLE IMAGING STUDIES WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. THEREFORE, WE ARE UNABLE TO DETERMINE THE DEFINITIVE CAUSE OF THE REPORTED EVENT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH AN INDICATION OF L4-S1 CORONAL DEFORMITY WITH FORAMENAL STENOSIS IN NEED OF L4-S1 PIF IN SPINAL THERAPY. IT WAS REPORTED THAT THE PATIENT'S BONE WAS EXTREMELY SCLEROTIC AND OVERGROWN, THEREFORE IT WAS VERY HARD AND REQUIRED ADDITIONAL FORCE TO BE USED FROM THE SURGEON, THE TIP OF THE INSTRUMENT BENT AT THIS POINT. THERE WAS A DELAY OF ABOUT 5 MINS REPORTED. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FRAGMENTS IN THE PATIENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1017534 TSRH 3DX PROBE HXB MEDTRONIC SOFAMOR DANEK USA, INC 8350294 KH18J868 00613994282668

Patients

Seq Age Sex Outcome Treatment
1