FDA Adverse Event Malfunction Summary report: N

LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL

MDR report key: 1211901 · Received October 23, 2008

Report

Report Number
3005075853-2008-02541
Event Type
Malfunction
Date Received
October 23, 2008
Date of Event
September 30, 2008
Report Date
October 2, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DROPPING OR EJECTED CLIP. EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RETURNED IN GOOD VISUAL CONDITION. THE INSTRUMENT WAS CYCLED AND EJECTED THE REMAINING CLIPS. THE INSTRUMENT LOCKED OUT AS INTENDED. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MFG PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO AN UNKNOWN PROCEDURE, JAMMING OCCURRED AND THE CLIP EJECTED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL FZP ETHICON ENDO-SURGERY, LLC NA UNK

Patients

Seq Age Sex Outcome Treatment
1