FDA Adverse Event
Malfunction
Summary report: N
LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL
MDR report key: 1211901
·
Received October 23, 2008
Report
- Report Number
- 3005075853-2008-02541
- Event Type
- Malfunction
- Date Received
- October 23, 2008
- Date of Event
- September 30, 2008
- Report Date
- October 2, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DROPPING OR EJECTED CLIP. EVAL SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RETURNED IN GOOD VISUAL CONDITION. THE INSTRUMENT WAS CYCLED AND EJECTED THE REMAINING CLIPS. THE INSTRUMENT LOCKED OUT AS INTENDED. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MFG PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT PRIOR TO AN UNKNOWN PROCEDURE, JAMMING OCCURRED AND THE CLIP EJECTED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL | FZP | ETHICON ENDO-SURGERY, LLC | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |