FDA Adverse Event
Malfunction
Summary report: N
TOALCARE
MDR report key: 1211867
·
Received October 22, 2008
Report
- Report Number
- 1824206-2008-03872
- Event Type
- Malfunction
- Date Received
- October 22, 2008
- Date of Event
- September 25, 2008
- Report Date
- September 25, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE GROUND PRONG MISSING FROM THE AC POWER CORD. THE TECHNICIAN REPLACED THE POWER CORD TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE BED HAS LOST GROUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOALCARE | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |