SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 1211329 · Received October 22, 2008

Report

Report Number
2023826-2008-01334
Event Type
Malfunction
Date Received
October 22, 2008
Date of Event
September 16, 2008
Report Date
September 25, 2008
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P880091
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: CONCLUSION: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CAN NOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES, AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED, AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED AN AA4204VL SILICONE PLATE LENS AND IT TORE AS IT WAS ADVANCING INTO THE EYE. THE LENS WAS REMOVED WITHOUT ANY PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL CO. AA4204VF NA

Patients

Seq Age Sex Outcome Treatment
1 UNK INJECTOR: MODEL MSI-TR| STAARVISC II: LOT# UNK| CARTRIDGE: MODEL MTC-60CFP