SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA
Report
- Report Number
- 2023826-2008-01318
- Event Type
- Malfunction
- Date Received
- October 22, 2008
- Date of Event
- August 12, 2008
- Report Date
- September 26, 2008
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P990048
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION RESULTS: (OTHER): VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWED THAT A TEAR IN THE OPTIC AND ONE OF THE HAPTICS WAS TORN. THERE WAS EVIDENCE OF A CLEAR SURGICAL RESIDUE. CONCLUSION: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. THE DAMAGE TO THE LENS, AS OBSERVED AND PHOTOGRAPHED UPON THE RETURN OF THE LENS TO STAAR, CAN NOT BE DEFINITIVELY AND EXCLUSIVELY CORRELATED TO A SPECIFIC ROOT CAUSE. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES, AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED, AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.
IT WAS REPORTED THAT THE SURGEON INSERTED A THREE PIECE SILICONE LENS AND THE LENS TORE WHILE BEING INSERTED. THERE WAS PT CONTACT, BUT NO INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | AQ2003V | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INJECTOR: MODEL MSI-PM| CARTRIDGE: MODEL AQ CARTRIDGE-FP |