FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1211109 · Received October 23, 2008

Report

Report Number
3003464075-2008-00511
Event Type
Injury
Date Received
October 23, 2008
Date of Event
September 25, 2008
Report Date
September 25, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
PMA / PMN Number
K050525
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK OF THE PATIENT'S BLOOD IN A TIMELY MANNER AS INSTRUCTED IN THE USER'S GUIDE. EVAL OF THE RETURNED CARTRIDGE IDENTIFIED A DIALYSATE LEAK IN THE DISPOSABLE FLUID PATH. THE OCCURRENCE OF SIMILAR LEAKS HAS BEEN INVESTIGATED AND AN EXACT ASSIGNABLE CAUSE CANNOT BE DETERMINED. APPROPRIATE ACTION HAS BEEN TAKEN TO REDUCE THE OCCURRENCE OF LEAKS. THE USER'S GUIDE INCLUDES ADEQUATE WARNINGS FOR THE OPERATOR TO PERIODICALLY CHECK THE SYSTEM FOR LEAKS AS THE CYCLER MAY NOT SENSE SLOW LEAKS AND TO TERMINATE THE TREATMENT IF A LEAK CANNOT BE RESOLVED. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFO WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A SYSTEM ALARM OCCURRED DURING A ROUTINE HEMODIALYSIS TREATMENT. A CLEAR FLUID LEAK WAS OBSERVED. THE OPERATOR DID NOT PERFORM RINSEBACK, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR - 170 8057704

Patients

Seq Age Sex Outcome Treatment
1 72 YR Other