FDA Adverse Event Malfunction Summary report: N

*

MDR report key: 1210885 · Received July 30, 2008

Report

Report Number
2250051-2008-00389
Event Type
Malfunction
Date Received
July 30, 2008
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS, INC.
Product Code
JTC
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SVC ENGINEER (FSE) EVALUATED THE INSTRUMENT AND REPLACED PLUNGER CLAMP POSITION #2. THE INSTRUMENT WAS TESTED WITHOUT FURTHER PROBLEM AND WAS RETURNED TO EXPECTED OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * * JTC ORTHO-CLINICAL DIAGNOSTICS, INC. * *

Patients

Seq Age Sex Outcome Treatment
1