DIMENSION VISTA 1500
Report
- Report Number
- 2517506-2021-00150
- Event Type
- Malfunction
- Date Received
- July 2, 2021
- Date of Event
- June 9, 2021
- Report Date
- August 12, 2021
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJE
- PMA / PMN Number
- K051087
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SIEMENS FILED THE INITIAL MDR 2517506-2021-00150 ON 02JUL2021. ADDITIONAL INFORMATION (21JUL2021): THE CUSTOMER CONTACTED THE SIEMENS CUSTOMER CARE CENTER (CCC) AND A SIEMENS CUSTOMER SERVICE SPECIALIST (CSS) REVIEWED THE INFORMATION PROVIDED. A REVIEW OF THE DIMENSION VISTA 1500 (SN: (B)(6)) INSTRUMENT DID NOT FIND ANY ANOMALIES OR ISSUES WITH THE INSTRUMENT. HOWEVER, A REVIEW OF THE ASPIRATION PROFILE ON DIMENSION VISTA 1500 (SN: (B)(6)) SHOWED THE UNDILUTED SAMPLE HAD AN ATYPICAL ASPIRATION PROFILE INDICATING A POTENTIAL SAMPLE ISSUE SUCH AS PARTIAL CLOTS, OR HIGHER SAMPLE VISCOSITY. THE CALCULATED VALUES FOR THE REACTIONS SHOWS THAT A FLEX SEQUENCE NUMBER HAD AN ATYPICALLY HIGH POLISHED MILLIABSORBANCE UNITS (MAU) FOR BOTH OF THE REACTION READ POINTS COMPARED TO ALL OTHER REAGENT FLEX CARTRIDGES. ALLIANCE APPLICATIONS REAGENT FLEX CARTRIDGES ARE EMPTY CARTRIDGES THAT ARE MANUALLY FILLED WITH THE ALLIANCE APPLICATION BULK REAGENT BY THE OPERATOR BEFORE LOADING ON THE ANALYZER. CONSULTATION WITH THE BULK REAGENT MANUFACTURER CONFIRMED THAT THIS INCREASE IN THE POLISHED MAU IS CONSISTENT WITH THE REAGENT 1 AND THE REAGENT 2 BEING SWITCHED IN THE REAGENT FLEX CARTRIDGE. THIS LED TO A CHANGE IN THE CURVE SHAPE WHERE SAMPLES ABOVE THE ASSAY RANGE COULD RECOVER WITHIN THE ASSAY RANGE. QUALITY CHECKS (QC) PERFORMED OUT OF THE AFFECTED REAGENT CARTRIDGE SHOWED A SHIFT DOWN IN RECOVERY COMPARED TO OTHER DATA POINTS BUT WAS WITHIN THE OPERATOR RANGES. SIEMENS FOUND THAT THIS WAS THE ONLY FLEX REAGENT CARTRIDGE TO DEMONSTRATE THIS ISSUE. SIEMENS CONFIRMED THAT THE CAUSE OF THE ISSUE IS THE INCORRECT FILLING OF THE SINGLE REAGENT FLEX CARTRIDGE LEADING TO REAGENT 1 AND REAGENT 2 TO BE SWAPPED. SYSTEM IS FULLY OPERATIONAL. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED. MDR 2517506-2021-00149_S1 WAS FILED IN CONJUNCTION TO THIS REPORT.
SIEMENS IS INVESTIGATING THE ISSUE. MDR 2517506-2021-00149 WAS FILED IN CONJUNCTION TO THIS REPORT FOR DIMENSION VISTA 1500 INSTRUMENT (SERIAL: (B)(4)).
MULTIPLE DISCORDANT, FALSELY DEPRESSED METHOTREXATE TEST RESULTS WERE OBTAINED ON A PATIENT SAMPLE, ON A DIMENSION VISTA 1500 (SERIAL: (B)(4)) INSTRUMENT. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN(S). THE SAME SAMPLE WAS REPEATED ON THE SAME INSTRUMENT AND AN ALTERNATE DIMENSION VISTA 1500 (SERIAL: SERIAL: (B)(4)) INSTRUMENT. THE SAMPLE WAS DILUTED 100-FOLD AND REPEATED ON THE ALTERNATE DIMENSION VISTA 1500 INSTRUMENT AND THE ORIGINAL INSTRUMENT. THE CUSTOMER STATED THAT THE PHYSICIAN(S) EXPECTED A MUCH HIGHER RESULT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE FALSELY DEPRESSED METHOTREXATE PATIENT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1007019 | DIMENSION VISTA 1500 | DIMENSION VISTA 1500 | JJE | SIEMENS HEALTHCARE DIAGNOSTICS INC. | DIMENSION VISTA 1500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |