FDA Adverse Event Injury Summary report: N

STERRAD NX STERILIZER

MDR report key: 1210747 · Received October 23, 2008

Report

Report Number
2084725-2008-00694
Event Type
Injury
Date Received
October 23, 2008
Date of Event
September 24, 2008
Report Date
September 24, 2008
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
MLR
PMA / PMN Number
K042116
Removal / Correction Number
Z-1631-2008
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESPIRATORY EMBARRASSMENT, BURNS ON THE LEVEL OF THE THROAT, IRRITATED EYES. OIL MIST. CAPITAL EQUIPMENT, EVALUATED AT CUSTOMER SITE. THE FSE PERFORMED THE FOLLOWING CHECKS AND TESTS: OIL LEVEL IN THE VACUUM PUMP, LEAK OF THE CHAMBER AND OIL FILTER AND VALVE. THREE EMPTY CYCLES WERE RUN WITHOUT WHITE SMOKE FROM OUTPUT OF THE OIL FILTER. THERE WAS NO WHITE SMOKE NOR SMELL. AS A CAUTION MEASURE, REPLACEMENT OF THE OIL FILTER. REFERENCE MDRS: 2084725-2008-00695, 00696, 00697, 00698, 00699, 00700, 00701, 00702 AND 00703.

Description of Event or Problem · 1

THE AFFILIATES REPORTED THAT THE CUSTOMER EXPERIENCED THE LOW AREA H2O2 CANCELLATION. THE CUSTOMER EDITED CYCLE TICKET, OPENED MACHINE DOOR, AND BROUGHT THE LOAD FROM THE STERILIZATION ROOM TO THE WASHING ROOM. THE CUSTOMER DID NOT MENTION THE PRESENCE OF SMOKE OR SMELL IN THE STERILIZATION ROOM WHEN OPENING THE DOOR MACHINE TO TAKE THE LOAD. ONE TO TWO MINUTES LATER, THE CUSTOMER NOTICED A SMELL IN THE STERILIZATION ROOM. THEY ALSO NOTICED A WHITE SMOKE IN THE WASHING ROOM BUT NOT IN THE STERILIZATION ROOM. THE CUSTOMER CONFIRMED THAT THEY DID NOT MODIFY WASHING PROCEDURE FOR CYSTOSCOPE NOR CHANGE DETERGENT. THE CUSTOMER REPORTED TEN EXPERIENCED "RESPIRATORY EMBARRASSMENT, BURNS ON THE LEVELS OF THROAT, HEADACHE AND EYE IRRITATION." ONE OF THE TEN PEOPLE WENT TO THE OPHTHALMOLOGIST FOR PERSISTENT EYE IRRITATION. IT WAS ALSO REPORTED THAT AFTER FIVE DAYS, THE SYMPTOMS OF THE TEN PEOPLE HAVE RESOLVED. THE FIELD SERVICE ENGINEER WENT TO THE FACILITY TO ASSESS THE UNIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERRAD NX STERILIZER STERRAD EQUIPMENT (MLR) MLR ADVANCED STERILIZATION PRODUCTS NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK STERRAD CASSETTE