FDA Adverse Event Malfunction Summary report: N

ALARIS 8015 PCU POINT OF CARE UNIT INFUSION PUMP 5.7" SCREEN

MDR report key: 12102070 · Received July 1, 2021

Report

Report Number
3007409280-2021-00008
Event Type
Malfunction
Date Received
July 1, 2021
Date of Event
June 14, 2021
Report Date
June 14, 2021
Manufacturer
CARDINAL HEALTH 303, INC.
Product Code
FRN
PMA / PMN Number
K091308
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS HEATING UP. THE BIOMED NOTICED THE UNIT HEATING UP NEAR THE ALUMINUM MOUNTING BRACKET AFTER BEING PLUGGED IN FOR A PROLONGED AMOUNT OF TIME. HE BELIEVES THE ISSUE IS NOT WITH THE BATTERY; RATHER, THE POWER SUPPLY BOARD. THERE WAS NO PATIENT INVOLVEMENT. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. TECHNICAL EVALUATION IDENTIFIED A BOARD MALFUNCTION; SPECIFICALLY, A DEFECTIVE SUPPLY POWER BOARD THAT WAS GETTING HOT. ADDITIONALLY, THE KEYPAD WAS SEPARATING FROM THE FRONT BEZEL. THE CUSTOMER COMPLAINT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE THE KEYPAD ADHESIVE WAS COMING UNDONE. THE 24V POWER SUPPLY AND THE POWER SUPPLY PCBA WERE REPLACED. THE DEVICE WAS REPAIRED AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF SERVICE AND TESTING. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS TYPE OF EVENT WILL CONTINUE TO BE MONITORED.

Description of Event or Problem · 1

THE DEVICE WAS HEATING UP. THE BIOMED NOTICED THE UNIT HEATING UP NEAR THE ALUMINUM MOUNTING BRACKET AFTER BEING PLUGGED IN FOR A PROLONGED AMOUNT OF TIME. HE BELIEVES THE ISSUE IS NOT WITH THE BATTERY; RATHER, THE POWER SUPPLY BOARD. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1001808 ALARIS 8015 PCU POINT OF CARE UNIT INFUSION PUMP 5.7" SCREEN ALARIS PC UNIT, MODELS 8000 AND 8015 FRN CARDINAL HEALTH 303, INC. FIAL2016

Patients

Seq Age Sex Outcome Treatment
1