ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER,4
Report
- Report Number
- 3005075853-2008-02463
- Event Type
- Malfunction
- Date Received
- October 20, 2008
- Date of Event
- September 3, 2008
- Report Date
- September 4, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B) (4). (B) (4). RESULTS & CONCLUSIONS - DAMAGED FIRING TRIGGER TEETH. H3. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH NO RELOAD PRESENT. NO FUNCTIONAL TEST COULD BE PERFORMED WITH THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. ALTHOUGH NO CONCLUSION COULD BE REACHED ON HOW THE DEVICE GOT DAMAGED WITH THE INFORMATION PROVIDED, IT SHOULD BE NOTED THAT AT LEAST A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B) (4). EVENT COULD NOT BE CONFIRMED AS NO RELOAD WAS RECEIVED FOR ANALYSIS. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN OVARIAN CYST PROCEDURE, THE DEVICE WOULD NOT FIRE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER,4 | GDW | ETHICON ENDO-SURGERY, LLC | NA | E4KU2M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |