FDA Adverse Event Malfunction Summary report: N

ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER,4

MDR report key: 1210140 · Received October 20, 2008

Report

Report Number
3005075853-2008-02463
Event Type
Malfunction
Date Received
October 20, 2008
Date of Event
September 3, 2008
Report Date
September 4, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B) (4). (B) (4). RESULTS & CONCLUSIONS - DAMAGED FIRING TRIGGER TEETH. H3. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AND WITH NO RELOAD PRESENT. NO FUNCTIONAL TEST COULD BE PERFORMED WITH THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH WERE FOUND BROKEN. ALTHOUGH NO CONCLUSION COULD BE REACHED ON HOW THE DEVICE GOT DAMAGED WITH THE INFORMATION PROVIDED, IT SHOULD BE NOTED THAT AT LEAST A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B) (4). EVENT COULD NOT BE CONFIRMED AS NO RELOAD WAS RECEIVED FOR ANALYSIS. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN OVARIAN CYST PROCEDURE, THE DEVICE WOULD NOT FIRE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ETS FLEX45 ENDOSCOPIC ARTICULATING LINEAR CUTTER,4 GDW ETHICON ENDO-SURGERY, LLC NA E4KU2M

Patients

Seq Age Sex Outcome Treatment
1