FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 1210126
·
Received October 20, 2008
Report
- Report Number
- 3002158293-2008-00550
- Event Type
- Malfunction
- Date Received
- October 20, 2008
- Date of Event
- October 1, 2008
- Report Date
- October 17, 2008
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK HAS BEEN COMPLETED. THE REPORTED PROBLEM WAS CONFIRMED. THE CAUSE OF THE BATTERY PACK NOT CHARGING WAS WATER WITHIN THE BATTERY PACK. THE ROOT CAUSE OF THE WATER IS NOT KNOWN, BUT IS PROBABLY DUE TO THE PT SUBMERSING THE BATTERY PACK IN WATER. THE BATTERY PACK WAS SCRAPPED. NO ADVERSE EVENT RESULTED FROM THE FAULTY BATTERY PACK. THE LAST PT TO USE THIS BATTERY PACK DID NOT REPORT ANY PROBLEMS.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING OF BATTERY PACK WHICH WAS RETURNED FOR ROUTINE MAINTENANCE, IT WAS DISCOVERED THAT THE BATTERY PACK WOULD NOT CHARGE. THE LAST PATIENT TO USE THIS BATTERY PACK DID NOT REPORT ANY PROBLEMS WITH IT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |