THERAPY ABLATION CATHETER
Report
- Report Number
- 2030404-2008-00015
- Event Type
- Malfunction
- Date Received
- October 17, 2008
- Date of Event
- September 25, 2008
- Report Date
- September 25, 2008
- Manufacturer
- IRVINE BIOMEDICAL, INC.
- Product Code
- LPB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
NO PT CONSEQUENCES WERE REPORTED. ONE USED CATHETER WAS RECEIVED ON 10/7/2008. THE UNIT WAS VISUALLY AND FUNCTIONALLY EXAMINED. RESULTS OF THE TESTING AREA AS FOLLOWS: VISUAL INSPECTION FOUND CATHETER SHAFT DAMAGE AT THE DISTAL FUSE AREA 7.0 CM FROM THE DISTAL TIP. IT WAS NOTED THAT THE SHAFT WAS TWISTED NEAR THE DAMAGED LOCATION WHICH SUGGESTS THE CATHETER WAS SUBJECTED TO OVER-TORQUE. THE CATHETER WAS, HOWEVER, HELD TOGETHER BY THE INTERNAL COMPONENTS AS DESIGNED. CATHETERS FROM SIMILAR TUBING SHAFTS WERE PULLED FROM FINISHED GOODS INVENTORY AND WORST CASE CONDITIONS WERE SIMULATED. THE CATHETERS WERE PRECONDITIONED FOR 4 HOURS IN A HEATED 37 DEGREE CELSIUS WATER BATH (BODY TEMPERATURE), SUBJECTED TO REPETITIVE CYCLES OF FLEXION/DEFLECTION (40 CYCLES STEERING) FOLLOWED BY 720 DEGREE TWISTING FORCE (TORQUE) BASED ON THE SAME ACCEPTANCE CRITERIA ESTABLISHED DURING PRODUCT DESIGN. WITH EACH ROTATION THE FUSE AREA WAS EXAMINED UNDER A MICROSCOPE. EVIDENCE OF STRESS BUILDUP WAS OBSERVED DURING THE SIMULATION TESTING, HOWEVER, ALL THREE CATHETERS RETAINED MECHANICAL INTEGRITY AT GREATER THAN 2X THE ACCEPTANCE CRITERIA. THE FAILURE MODE WAS NOT DUPLICATED EVEN AFTER 10 FULL ROTATIONS IN THE SAME DIRECTION WHICH IS BEYOND NORMAL USE. THE FUSING PROCESS IS CLOSELY MONITORED. PROCESS MONITORING DATA WAS REVIEWED FOR THE PERIOD OF MANUFACTURE OF THIS LOT AND EXCEEDS MINIMUM TENSILE STRENGTH REQUIREMENTS OF 3.37 POUNDS PER ISO. IN ALL CASES, THE CATHETER SHAFT STRETCHED BEFORE A FUSE JOINT FAILURE OCCURRED, INDICATING A STRONG FUSE JOINT. DEVICE HISTORY RECORD FOR LOT K04540 WAS ALSO REVIEWED TO ENSURE ALL MANUFACTURING AND INSPECTION PROCEDURES WERE PERFORMED AND REPORTED AS COMPLETE AND ACCEPTABLE. NO EVIDENCE OF A MANUFACTURING ERROR COULD BE FOUND. YEARS 2007 AND 2008 COMPLAINTS WERE REVIEWED. THIS IS A RARE OCCURRENCE. THE TEST DATA SUGGESTS THAT THE CATHETER WAS OVER-TORQUED BEYOND NORMAL USE. SINCE LOAD OR TENSION MUST BE PLACED ONTO THE CATHETER, THE DISTAL SECTION MAY HAVE BEEN ENTRAPPED IN A POCKET OF TISSUE. THE FUSING PROCESS AND TUBING TENSILE STRENGTH HAVE BEEN VALIDATED TO ENSURE MECHANICAL INTEGRITY UNDER NORMAL USE CONDITIONS. THE INSTRUCTION FOR USE STATE "EXCESSIVE BENDING OR KINKING OF THE CATHETER MAY CAUSE DAMAGE TO THE CATHETER."
DAMAGE TO THE DISTAL SECTION OF THE CATHETER SHAFT WAS REPORTED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERAPY ABLATION CATHETER | ABLATION CATHETER | LPB | IRVINE BIOMEDICAL, INC. | 83352 | K04540 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |