PUMP MMT-1782K 670G V4.11 MM
Report
- Report Number
- 2032227-2021-163156
- Event Type
- Malfunction
- Date Received
- July 1, 2021
- Date of Event
- June 25, 2021
- Report Date
- May 13, 2022
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000264727
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
RETAINER RING = BLACK. CUSTOMER RETURNED INSULIN PUMP FOR AN ALLEGED BLANK DISPLAY FOUND ON 06/25/2021. DEVICE P-CAP / TEST RESERVOIR LOCKS IN PLACE PROPERLY. DEVICE POWERED UP PROPERLY AFTER BATTERY INSTALLATION. NO BLANK DISPLAY NOTED. DEVICE PASSED THE DISPLACEMENT TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT AND SELF TEST. DEVICE WAS MONITORED AND NO UNEXPECTED POWERING OFF OR BLANK DISPLAY NOTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS SOFTWARE. HOWEVER, BATT OUT LIMIT(6)-06/25/2021 06:28:23.000, CHANGE BATTERY FAULT (73)-06/25/2021 05:46:00.000 WERE FOUND IN HISTORY FILE. POWER MANAGEMENT TOOL CONFIRMED UNLOADED VLITH VOLTAGE AND LOADED VLITH VOLTAGE IS WITHIN SPECIFICATIONS. DEVICE WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE PCBA1, PCBA2, FORCE SENSOR, MOTOR AND BATTERY TUBE ASSEMBLY NOTED. DEVICE WAS RECEIVED WITH SCRATCHED LCD WINDOW. IN SUMMARY, CUSTOMER ALLEGED BLANK DISPLAY NOT CONFIRMED. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND THE DISPLAY DID NOT RETURN. CUSTOMER STATED THAT THE DISPLAY WAS NOT BLANK LESS THAN 30 SECONDS AND DID NOT RETURN. CUSTOMER STATED THAT THE CONTACTS ON THE BATTERY CAP WERE NOT MISSING OR DAMAGED. CUSTOMER STATED THAT THE BATTERY COMPARTMENT WAS NEITHER DAMAGED NOR CORRODED ALSO THE SPRING WAS NEITHER DAMAGED NOR CORRODED. THE DISPLAY DID NOT RETURN AFTER INSULIN PUMP RESTART. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1001923 | PUMP MMT-1782K 670G V4.11 MM | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1782K | HG5163B | 000000763000264727 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |