FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY

MDR report key: 12096355 · Received June 30, 2021

Report

Report Number
3013756811-2021-69091
Event Type
Malfunction
Date Received
June 30, 2021
Date of Event
June 11, 2021
Report Date
June 30, 2021
Manufacturer
TANDEM DIABETES CARE
Product Code
OZO
UDI-DI
00853052007325
PMA / PMN Number
K201214
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE FAILURE INVESTIGATION HAS BEEN COMPLETED. BASED ON THE ANALYSIS, THE ALLEGED ISSUE WAS VERIFIED, AND A DIFFERENT ISSUE WAS IDENTIFIED (USB COMMUNICATION INOPERABLE).

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A CONTINUOUS GLUCOSE MONITOR ERROR 42. ADDITIONALLY, IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET. REPORTEDLY, THE ALERT CLEARED AND THE PUMP SUCCESSFULLY BEGAN CHARGING AFTER LEAVING THE PUMP PLUGGED INTO THE POWER SOURCE. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS BETWEEN 130-185. REPORTEDLY, THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
987735 T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO TANDEM DIABETES CARE 1000354 00853052007325

Patients

Seq Age Sex Outcome Treatment
1 65 YR INFUSION SET: AUTOSOFT XCINSULIN: NOVOLOG| INFUSION SET: AUTOSOFT XCINSULIN: NOVOLOG