FDA Adverse Event Malfunction Summary report: N

VERTEX RECONSTRUCTION SYSTEM

MDR report key: 12092289 · Received June 30, 2021

Report

Report Number
1030489-2021-00870
Event Type
Malfunction
Date Received
June 30, 2021
Date of Event
June 9, 2021
Report Date
December 23, 2021
Manufacturer
MDT SOFAMOR DANEK PUERTO RICO MFG
Product Code
KWP
PMA / PMN Number
SEE H10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2021, MEDICAL ADVISOR INTERPRETATION: IMAGES SHOW C5 SCREW FRACTURE. THE BROKEN FRAGMENT REMAIN IN PATIENT. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS #271043075:G6956430 LOT# H5389079 OPTICAL EXAMINATION OF THE ADJACENT SURFACE IMMEDIATELY AROUND THE CRACK ORIGINATION AREA DID NOT IDENTIFY A SURFACE DEFECT WHICH COULD CONTRIBUTE TO CRACK PROPAGATION. MICROSCOPIC EXAMINATION OF FRACTURE SURFACES REVEALS A FAIRLY FLAT, QUASI-BRITTLE FRACTURE, WITH PROGRESSIVE STRIATIONS, WHICH ARE INDICATIVE OF CYCLIC FATIGUE, FOLLOWED BY ULTIMATE FAILURE OF THE IMPLANT. THE FRACTURE SURFACE MORPHOLOGY SUGGESTS THE PRIMARY MODE OF FRACTURE TO BE CYCLIC FATIGUE, WITH A LOCALIZED REGION OF ORIGIN, DIRECTION OF PROPAGATION AND SUBSEQUENT FINAL FRACTURE. THIS IS CONSISTENT WITH CYCLIC FATIGUE. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS PART IS NOT APPROVED FOR USE IN THE UNITED STATES; HOWEVER A LIKE DEVICE CATALOG # 6958712, 510K # K042789, UDI # (B)(4) WAS CLEARED IN THE UNITED STATES. THUS, IT IS REPORTABLE FOR MALFUNCTION, NOT SERIOUS INJURY, ALTHOUGH SURGICAL INTERVENTION DID TAKE PLACE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH A SCREW. IT WAS REPORTED THAT  ON (B)(6) 2020 PATIENT WAS IMPLANTED WITH A SCREW. AS THE SCREW WAS BROKEN THE PATIENT HAD TO UNDERGO REVISION SURGERY FOR  C5L REMOVAL, C5R REPLACEMENT, FUSION EXTENSION. THE BONES WERE NOT FUSED. THE BROKEN PART OF THE SCREW WAS NOT A PART THAT COULD BE REMOVED, SO IT REMAINED. PATIENT WAS HOSPITALIZED BECAUSE IT WOULD BE THE SAME AS THE UNFIXED STATE DUE TO BREAKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
992261 VERTEX RECONSTRUCTION SYSTEM APPLIANCE, FIXATION, SPINAL INTERLAMINAL KWP MDT SOFAMOR DANEK PUERTO RICO MFG G6956430 H5389079

Patients

Seq Age Sex Outcome Treatment
1 65 YR Male Hospitalization| R