FDA Adverse Event Malfunction Summary report: N

ROSA ONE

MDR report key: 12090620 · Received June 30, 2021

Report

Report Number
3009185973-2021-00159
Event Type
Malfunction
Date Received
June 30, 2021
Date of Event
June 2, 2021
Report Date
January 6, 2022
Manufacturer
MEDTECH SAS
Product Code
HAW
PMA / PMN Number
K200511
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A FULL ANALYSIS OF THE DATA LOGS HAS BEEN PERFORMED AND THIS ANALYSIS CONCLUDED THAT THE PATIENT FOLDER EXPORTED ON THE USB WAS CORRUPTED. THIS CAUSED A WINDOWS ERROR EACH TIME THE PATIENT FOLDER WAS ATTEMPTED TO BE OPENED. HOWEVER, THE CAUSE FOR THE CORRUPTION COULD NOT BE FOUND.

Description of Event or Problem · 0

IT WAS REPORTED THAT SURGEON TRIED TO LOAD IN PATIENT FOLDER FROM PLANNING STATION, AND WAS UNABLE TO LOAD. WHEN PATIENT FOLDER WAS ATTEMPTED TO BE LOADED, WINDOWS BLUE SCREEN ERROR OCCURRED, WITH MESSAGE 'PAGE FAULT IN NONPAGED AREA'. THE CLINICAL REPRESENTATIVE (CR) ARRIVED ON SITE, AND ATTEMPTED TO TROUBLESHOOT PATIENT FOLDER ISSUES, AND WAS UNABLE TO FIND .ROS FILE IN PATIENT FOLDER. SURGEON NEEDED TO RELOAD PATIENT SERIES AND RE-PLAN PROCEDURE BECAUSE PLANNING LAPTOP WAS UNAVAILABLE AND AT OTHER HOSPITAL LOCATION. CR ATTEMPTED TO TROUBLESHOOT LOADING OF SERIES FROM USB AND WAS UNABLE TO LOAD SERIES FROM FOLDER WHICH CONTAINED MRI POST-OPERATIVE. SOME OF SLICES OF SERIES COULD BE UN-ENCRYPTED, BUT COULD NOT UN-ENCRYPT MAJORITY OF SERIES, AND COULD THEREFORE NOT TROUBLESHOOT FULL SERIES. CR COULD NOT OPEN ON BUSINESS LAPTOP OR ROSA ROBOT, AND WAS NOT ABLE TO VIEW IN ROSANNA SOFTWARE, IQVIEW, OR SANTE DICOM EDITOR SOFTWARE WITHOUT CAUSING PAGE FAULT ERROR. AS A RESULT, IT WAS NECESSARY TO PULL AFFECTED DICOM SERIES ON USB FROM FIELD. DELAY OF 2 HOURS FOR REPLANNING, NO PATIENT INVOLVEMENT, PATIENT WAS NOT IN OR UNDER ANESTHESIA, BUT DELAY OF PROCEDURE START, PATIENT HAD TO WAIT IN PREOPERATIVE CARE UNTIL SURGERY WAS RE-PLANNED. SURGEON REPORTED NO ISSUES OR ERRORS UPON EXPORTING THE PATIENT FOLDER.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN EVALUATED YET FOR INVESTIGATION PURPOSE. ONCE THE EVALUATION IS PERFORMED, A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED. UNIQUE IDENTIFIER (UDI) #: UNKNOWN.

Description of Event or Problem · 1

IT WAS REPORTED THAT SURGEON TRIED TO LOAD IN PATIENT FOLDER FROM PLANNING STATION, AND WAS UNABLE TO LOAD. WHEN PATIENT FOLDER WAS ATTEMPTED TO BE LOADED, WINDOWS BLUE SCREEN ERROR OCCURRED, WITH MESSAGE 'PAGE FAULT IN NONPAGED AREA'. THE CLINICAL REPRESENTATIVE (CR) ARRIVED ON SITE, AND ATTEMPTED TO TROUBLESHOOT PATIENT FOLDER ISSUES, AND WAS UNABLE TO FIND .ROS FILE IN PATIENT FOLDER. SURGEON NEEDED TO RE-LOAD PATIENT SERIES AND RE-PLAN PROCEDURE BECAUSE PLANNING LAPTOP WAS UNAVAILABLE AND AT OTHER HOSPITAL LOCATION. CR ATTEMPTED TO TROUBLESHOOT LOADING OF SERIES FROM USB AND WAS UNABLE TO LOAD SERIES FROM FOLDER WHICH CONTAINED MRI POST-OPERATIVE. SOME OF SLICES OF SERIES COULD BE UN-ENCRYPTED, BUT COULD NOT UN-ENCRYPT MAJORITY OF SERIES, AND COULD THEREFORE NOT TROUBLESHOOT FULL SERIES. CR COULD NOT OPEN ON BUSINESS LAPTOP OR ROSA ROBOT, AND WAS NOT ABLE TO VIEW IN ROSANNA SOFTWARE, IQVIEW, OR SANTE DICOM EDITOR SOFTWARE WITHOUT CAUSING PAGE FAULT ERROR. AS A RESULT, IT WAS NECESSARY TO PULL AFFECTED DICOM SERIES ON USB FROM FIELD. DELAY OF 2 HOURS FOR REPLANNING, NO PATIENT INVOLVEMENT, PATIENT WAS NOT IN OR UNDER ANESTHESIA, BUT DELAY OF PROCEDURE START, PATIENT HAD TO WAIT IN PREOPERATIVE CARE UNTIL SURGERY WAS RE-PLANNED. SURGEON REPORTED NO ISSUES OR ERRORS UPON EXPORTING THE PATIENT FOLDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
993321 ROSA ONE COMPUTER-ASSISTED SURGICAL DEVICE HAW MEDTECH SAS PLANNING STATION 3.1.4.1650

Patients

Seq Age Sex Outcome Treatment
1 Unknown