FDA Adverse Event Malfunction Summary report: N

SINGLE USE SPHINCTEROTOME V (DISTAL WIREGUIDED)

MDR report key: 12089960 · Received June 30, 2021

Report

Report Number
8010047-2021-08167
Event Type
Malfunction
Date Received
June 30, 2021
Date of Event
June 2, 2021
Report Date
June 30, 2021
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
KNS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR INVESTIGATION. THE CUTTING WIRE WAS BROKEN. THE BROKEN PORTION WAS SCORCHED AND MELTED. THE OUTER DIAMETER OF THE CUTTING WIRE WAS MEASURED. THE RESULT INDICATED NO ABNORMALITIES. THE LENGTH OF THE COATED PORTION OF THE CUTTING WIRE, AND THE CUTTING WIRE ITSELF PRESENTED NO ABNORMALITIES. THERE WERE NO MISSING PARTS IN THE SUBJECT DEVICE. OTHER ABNORMALITIES THAT COULD LEAD TO THE BREAKAGE OF THE CUTTING WIRE WERE NOT CONFIRMED. NO ABNORMALITIES WERE DETECTED IN THE DEVICE HISTORY RECORD WITH THE LOT NUMBER FOR THE FOLLOWING INSPECTION ITEMS WHICH RELATED TO THE REPORTED PHENOMENON. - LENGTH OF CUTTING WIRE - LENGTH OF COATED PORTION - OPERATION OF CUTTING WIRE BASED ON THE RESULTS OF CONFIRMATION OF THE DEVICE AND THE INVESTIGATION RESULTS IN THE PAST, A LIKELY MECHANISM CAUSING THE BROKEN CUTTING WIRE MIGHT BE THE FOLLOWING. THE DEVICE WAS NOT PROTRUDED ENOUGH FROM THE ENDOSCOPE UNTIL THE REAR END OF THE CUTTING WIRE WAS IN THE FIELD OF VIEW. DUE TO THE SITUATION OF ¿1¿ DESCRIPTION, THE CUTTING WIRE AND THE ENDOSCOPE WERE BEING CLOSE TO EACH OTHER. THE OUTPUT WAS ACTIVATED IN STATE OF ¿2¿ DESCRIPTION. THIS MIGHT HAVE LED TO AN ELECTRICAL DISCHARGE BETWEEN THE CUTTING WIRE AND THE DISTAL END OF THE ENDOSCOPE. AN ELECTRICAL DISCHARGE POSSIBLY OCCURRED, AND THE CUTTING WIRE BECAME HOT INSTANTLY. THAT MIGHT HAVE CAUSED THE CUTTING WIRE TO BREAK. THE ABOVE DEVICE HANDLING HAS WARNED IN THE INSTRUCTION MANUAL.

Description of Event or Problem · 0

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED BY A BIOMEDICAL ENGINEER THAT DURING AN ENDOSCOPIC SPHINCTEROTOMY USING THE SUBJECT DEVICE, THE KNIFE WIRE WAS BROKEN OFF ON ACTIVATION. THE SURGEON CHANGED ANOTHER DEVICE, BUT THE KNIFE WIRE WAS BROKEN OFF, TOO. IN ADDITION, THE SURGEON CHANGED, BUT THE KNIFE WIRE WAS BROKEN OFF, TOO. THE INTENDED PROCEDURE WAS COMPLETED WITH THE OTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THIS IS THE REPORT REGARDING THE 2ND BROKEN KNIFE WIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
993580 SINGLE USE SPHINCTEROTOME V (DISTAL WIREGUIDED) SINGLE USE PRELOADED SPHINCTEROTOME KNS OLYMPUS MEDICAL SYSTEMS CORP. KD-VC411Q-0725 11K

Patients

Seq Age Sex Outcome Treatment
1