FS DISPOSABLE INTERFACE
Report
- Report Number
- 3006695864-2021-08002
- Event Type
- Malfunction
- Date Received
- June 29, 2021
- Date of Event
- April 15, 2021
- Report Date
- August 16, 2021
- Manufacturer
- AMO MANUFACTURING USA, LLC
- Product Code
- GEX
- UDI-DI
- 15050474534688
- PMA / PMN Number
- K060372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES, RETURNED TO MANUFACTURER ON 07/22/2021. SECTION H3: DEVICE EVALUATED BY MANUFACTURER: YES . DEVICE EVALUATION: ONE (1) DEVICE WAS RETURNED WITHIN ITS ORIGINAL PACKAGING CONFIRMING THE REPORTED LOT NUMBER. A VISUAL INSPECTION OF THE RETURNED PRODUCT DID NOT REVEAL ANY DEBRIS, LOOSE PARTICLES, OR FIBERS. THE REPORTED ISSUE COULD NOT BE CONFIRMED. MANUFACTURING RECORD REVIEW: PER MANUFACTURING RECORDS REVIEW REPORT, NO RELATED DEVIATION, NCMR, NC OR CAPA WAS INITIATED DURING THE MANUFACTURING PROCESS OF THE REPORTED LOT#. ALL DEVICES MEET MATERIAL, ASSEMBLY, AND PERFORMANCE SPECIFICATIONS AT THE TIME OF PRODUCT RELEASE. CONCLUSION: BASED ON THE INVESTIGATION RESULTS THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
AGE/DATE OF BIRTH, WEIGHT, ETHNICITY: UNKNOWN/NOT PROVIDED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT THERE WAS DEBRIS ON THE CONE OF THE PATIENT INTERFACE THAT WAS BEING USED FOR THE OPERATIVE EYE. CONE HAD PATIENT CONTACT AND PROCEDURE WAS FIRED AND LOST WHILE THE CONE HAD DEBRIS. THERE WAS NO PATIENT HARM AND NO MEDICAL INTERVENTION NEEDED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 979426 | FS DISPOSABLE INTERFACE | POWERED LASER SURGICAL INSTRUMENT | GEX | AMO MANUFACTURING USA, LLC | 590106AN | 60278656 | 15050474534688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FEMTO SN (B)(6) |