FDA Adverse Event Malfunction Summary report: N

HOL ML 5MM ENDO APPLIER

MDR report key: 12085156 · Received June 29, 2021

Report

Report Number
3011137372-2021-00163
Event Type
Malfunction
Date Received
June 29, 2021
Date of Event
June 18, 2021
Report Date
June 18, 2021
Manufacturer
TELEFLEX MEDICAL
Product Code
GDO
UDI-DI
24026704696813
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THE DHR FOR THE RETURNED INSTRUMENT WAS REVIEWED AND FOUND COMPLETELY WITHOUT ANY NONCONFORMANCE'S OR IRREGULARITIES. IT WAS ALSO FOUND THAT THIS ORDER WAS MADE FROM THE CORRECT MATERIALS AND COMPONENTS. THIS INSTRUMENT WAS PRODUCED ON SHOP ORDER 50289610 AT THE TECOMET, INC. KENOSHA WI FACILITY AS PART OF A 50-PIECE LOT IN MARCH OF 2020. EVALUATION OF THE RETURNED INSTRUMENT SHOWS THAT THE JAWS ARE LOOSE AND MISALIGNED , AND THE JAW PIVOT PIN IS PULLED THRU ONE SIDE OF THE BENT/DAMAGED OUTER TUBE ASSEMBLY AND THE DRIVE ROD (N00185) IS ALSO BENT/DAMAGED IN THE AREA THAT IT ENGAGES THE JAWS. WE ARE ABLE TO VALIDATE THIS COMPLAINT. AFTER THE INITIAL EVALUATION THIS INSTRUMENT WAS DIS-ASSEMBLED IN ORDER TO EVALUATE ITS INTERNAL COMPONENTS AND IT WAS FOUND THAT THE INNER DRIVE ROD (N00185) FINGERS ARE BOTH DAMAGED WHERE THEY ENGAGE THE JAWS. MISHANDLING OF THIS DEVICE AT THE END USER'S FACILITY IS SUSPECTED. ALL 50 INSTRUMENTS FROM THIS LOT WERE 100% VISUALLY INSPECTED AND FUNCTION TESTED PRIOR TO SHIPMENT TO THE CUSTOMER AS THIS IS A STANDARDIZED PROCEDURE AT THIS FACILITY FOR THIS PRODUCT LINE. DUE TO THESE FINDINGS , NO FURTHER ACTIONS WILL BE TAKEN IN RESPONSE TO THIS COMPLAINT AND THIS RECORD WILL BE DEEMED CLOSED.

Description of Event or Problem · 0

THE PIVOT PIN GOT DETACHED DURING A SURGERY. THEREFORE, THE APPLIER WAS REPLACED WITH A NEW ONE. NOTHING FELL/REMAINED IN THE PATIENT. THE APPLIER WAS PURCHASED BY THE HOSPITAL IN (B)(6) 2021.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS NOT BEEN RETURNED FOR INVESTIGATION. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATED EVENTS.

Description of Event or Problem · 1

THE PIVOT PIN GOT DETACHED DURING A SURGERY. THEREFORE, THE APPLIER WAS REPLACED WITH A NEW ONE. NOTHING FELL/REMAINED IN THE PATIENT. THE APPLIER WAS PURCHASED BY THE HOSPITAL IN (B)(6) 2021.