FDA Adverse Event Malfunction Summary report: N

I-STAT CREA CARTRIDGE

MDR report key: 12084227 · Received June 29, 2021

Report

Report Number
2245578-2021-00057
Event Type
Malfunction
Date Received
June 29, 2021
Date of Event
June 22, 2021
Report Date
August 4, 2021
Manufacturer
ABBOTT POINT OF CARE
Product Code
CGL
UDI-DI
10054749000125
PMA / PMN Number
K973292
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 26-JUL-2021. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE LOT PASSED FINISHED GOODS RELEASE CRITERIA. RETAINED AND RETURNED CARTRIDGE TESTING MET THE APPLICABLE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AG, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR CREA CARTRIDGE LOT A21134.

Description of Event or Problem · 0

NA.

Additional Manufacturer Narrative · 1

APOC INCIDENT# (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON 25-JUN-2021, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CREA CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT CREATININE RESULT OF 2.9 ON AN 84 YEAR OLD FEMALE PATIENT PRESENTED FOR A STUDY SCAN AND ANGIOGRAM. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD:I-STAT ; RESULT : 2.9; DATE : (B)(6) 2021; TIME: 15:00. LAB , 0.9 , (B)(6) 2021 , 23:32. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
983837 I-STAT CREA CARTRIDGE CREA+ CARTRIDGE CGL ABBOTT POINT OF CARE NA A21134 10054749000125

Patients

Seq Age Sex Outcome Treatment
1 84 YR