I-STAT CREA CARTRIDGE
Report
- Report Number
- 2245578-2021-00057
- Event Type
- Malfunction
- Date Received
- June 29, 2021
- Date of Event
- June 22, 2021
- Report Date
- August 4, 2021
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- CGL
- UDI-DI
- 10054749000125
- PMA / PMN Number
- K973292
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 26-JUL-2021. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE LOT PASSED FINISHED GOODS RELEASE CRITERIA. RETAINED AND RETURNED CARTRIDGE TESTING MET THE APPLICABLE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. AG, APPENDIX 1- PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR CREA CARTRIDGE LOT A21134.
NA.
APOC INCIDENT# (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON 25-JUN-2021, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CREA CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT CREATININE RESULT OF 2.9 ON AN 84 YEAR OLD FEMALE PATIENT PRESENTED FOR A STUDY SCAN AND ANGIOGRAM. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD:I-STAT ; RESULT : 2.9; DATE : (B)(6) 2021; TIME: 15:00. LAB , 0.9 , (B)(6) 2021 , 23:32. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 983837 | I-STAT CREA CARTRIDGE | CREA+ CARTRIDGE | CGL | ABBOTT POINT OF CARE | NA | A21134 | 10054749000125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR |