Report

Report Number
8030673-2021-00177
Event Type
Malfunction
Date Received
June 28, 2021
Date of Event
June 1, 2021
Report Date
June 1, 2021
Product Code
CAI
UDI-DI
10190752147379
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: ONE OPENED SAMPLE PART NUMBER R6860K WAS RECEIVED FOR THE INVESTIGATION. BASED ON THE INVESTIGATION AND PER THE PHYSICAL SAMPLE AND PICTURES PROVIDED BY THE CUSTOMER, WE COULD OBSERVE THAT THE BLUE VALVE IS LOOSE INSIDE THE CUSHION, SO THE DEFECT WAS CONFIRMED. WE DETERMINED THAT THE MATERIAL COULD BE RELATED SINCE THE PART NUMBER R6860K IS NOT MANUFACTURED IN VYAIRE MEXICALI AND WE DO NOT MANIPULATE THE BLUE VALVE. IN ADDITION, THE PFMEA 76A REFERS TO THE PFMEA 83A FOR RAW MATERIAL, AND WE HAVE THE CONTROLS FOR THIS KIND OF ISSUE. AS A CORRECTIVE AND PREVENTIVE ACTION, SCAR MX-2021-030 WAS INITIATED FOR THE EXTERNAL SUPPLIER. ALSO, PERSONNEL WAS NOTIFIED ABOUT THE COMPLAINT.

Additional Manufacturer Narrative · 1

AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. NO ROOT CAUSE HAS BEEN DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE PRODUCT IS AVAILABLE TO BE RETURNED FOR INVESTIGATION. PICTURES ARE ALSO ATTACHED. THE CUSTOMER REPORTED TO VYAIRE MEDICAL THAT THE 6861- VITAL SIGNS® EXTRA LARGE FACE MASK WITH ADJUSTABLE AIR CUSHION SIZE 6X -LARGE ADULT FACE MASK, DEFLATED WHILE CONNECTED TO A PATIENT. THE CUSTOMER CONFIRMED THAT THERE WAS NO HARM AND THERE WAS NO INTERVENTION NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
977043 VITAL SIGNS® EXTRA LARGE FACE MASK WITH ADJUSTABLE AIR CUSHION SIZE 6X CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE) CAI VITAL SIGNS® EXTRA LARGE FACE MASK WITH ADJUSTABLE AIR CUSHION SIZE 6X -LARGE AD UNKNOWN 10190752147379

Patients

Seq Age Sex Outcome Treatment
1