FDA Adverse Event Malfunction Summary report: N

INOMAX DSIR PLUS (DELIVERY SYSTEM)

MDR report key: 12081072 · Received June 28, 2021

Report

Report Number
3004531588-2021-00155
Event Type
Malfunction
Date Received
June 28, 2021
Date of Event
December 17, 2020
Report Date
June 28, 2021
Manufacturer
MALLINCKRODT MANUFACTURING LLC.
Product Code
MRN
UDI-DI
00867538000106
PMA / PMN Number
K061901
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

REPORTABLE MALFUNCTION WITH INOMAX DSIR PLUS (B)(4) WAS DOCUMENTED AND INVESTIGATED UNDER (B)(4). A REVIEW OF THE DEVICE'S SERVICE LOG REVEALED THAT A DELIVERY FAILURE ALARM OCCURRED DUE TO AN APPARENT INTER-PROCESSOR LINK (IPL) FAILURE BETWEEN THE DEVICE'S MONITORING PROCESSOR AND DELIVERY PROCESSOR. ALTHOUGH IDENTIFIED IN THE DEVICE'S SERVICE LOG, THE REGIONAL SERVICE CENTER (RSC) DID NOT EXPERIENCE A DELIVERY FAILURE ALARM DURING TESTING. THE ROOT CAUSE FOR THIS OCCURRENCE WAS LIKELY DUE TO A MALFUNCTIONING PCA MAIN BOARD. AS A RESULT, THE DEVICE'S PCA MAIN BOARD WAS REPLACED. TRENDS WERE REVIEWED FOR THIS REPORTABLE CONDITION AND DETERMINED TO BE WITHIN THE ESTABLISHED CONTROL LIMITS. NO FURTHER ACTION IS REQUIRED AT THIS TIME. A FULL FUNCTIONAL TEST WAS PERFORMED AND THE DEVICE OPERATED ACCORDING TO SPECIFICATIONS. AS A RESULT, THE DEVICE WAS PLACED BACK INTO CIRCULATION FOR CUSTOMER USE.

Description of Event or Problem · 1

A DELIVERY FAILURE ALARM DUE TO AN APPARENT INTER-PROCESSOR LINK (IPL) FAILURE WAS IDENTIFIED BY A MALLINCKRODT EMPLOYEE DURING A ROUTINE SERVICE LOG REVIEW. THIS SERVICE LOG FINDING MEETS THE CRITERIA OF A REPORTABLE MALFUNCTION AS IT WAS LOCATED AT A USER'S FACILITY AND SET TO ADMINISTER A DOSE OF INHALED NITRIC OXIDE (NO) AT THE TIME OF THE EVENT. THIS DEVICE WAS LOCATED IN THE UNITED STATES WHEN THE REPORTABLE MALFUNCTION OCCURRED. THERE WAS NO DELIVERY FAILURE ALARM COMPLAINT OR REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
977297 INOMAX DSIR PLUS (DELIVERY SYSTEM) APPARATUS MRN MALLINCKRODT MANUFACTURING LLC. 10007 00867538000106

Patients

Seq Age Sex Outcome Treatment
1