PLATE COLUMBIA AG 5PRCT SB 90MM
Report
- Report Number
- 9680577-2021-00022
- Event Type
- Malfunction
- Date Received
- June 28, 2021
- Date of Event
- May 31, 2021
- Report Date
- October 21, 2021
- Manufacturer
- BECTON DICKINSON GMBH
- Product Code
- KZI
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6. INVESTIGATION: THIS STATEMENT IS TO SUMMARIZE FINDINGS ON THE RECENT COMPLAINT AGAINST COLUMBIA AGAR WITH 5% SHEEP BLOOD, CATALOG NUMBER 254071, LOT NUMBER 1082404. EVENT DESCRIPTION: IT WAS REPORTED THAT 7 PLATES WERE FOUND TO BE CONTAMINATED. COMPLAINT HISTORY REVIEW: THE COMPLAINT TRENDS WERE REVIEWED. THERE WERE NO SIMILAR COMPLAINTS REPORTED DURING THAT PERIOD ON THIS LOT NUMBER. HOWEVER, A TREND COULD NOT BE IDENTIFIED. BATCH HISTORY RECORD (BHR) REVIEW: THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY, AND NO DEVIATIONS WERE OBSERVED. SAMPLE ANALYSIS: PICTURE SAMPLE WAS PROVIDED SHOWING PLATES WITH CONTAMINATION. A VISUAL INSPECTION WAS PERFORMED ON THE RETENTION SAMPLE AND ONE STACK OF PLATES WERE INCUBATED. AS A RESULT OF THIS ANALYSIS NO CONTAMINATION WAS OBSERVED. EVALUATION RESULTS: AT THIS STAGE OF OUR INVESTIGATION, WE HAVE EXCLUDED ANY SYSTEMIC FAILURE IN OUR MANUFACTURING PROCESS. PLEASE NOTE THAT THIS PRODUCT DOES NOT HAVE AN SAL CLAIM. IT IS FILLED ASEPTICALLY, THEREFORE AN OCCURRENCE OF A CONTAMINATION EVENT CANNOT BE ENTIRELY RULED OUT. PLEASE NOTE THAT THE VALID STANDARD FOR THESE PRODUCTS IS DIN EN 12322 "IN VITRO DIAGNOSTIC MEDICAL DEVICES - CULTURE MEDIA FOR MICROBIOLOGY". ACCORDING TO THIS STANDARD THE CONTAMINATION RATE FOR EACH PRODUCT LOT MUST NOT EXCEED 4.9 %. FOR OUR CONTINUOUS MONITORING, WE DERIVE A CONTAMINATION RATIO FOR BELOW THIS SPECIFIED VALUE. ACCORDING TO OUR HIGH QUALITY STANDARD, WE ONLY RELEASE PRODUCT BATCHES TO THE MARKET WITH AN AQL (ACCEPTABLE QUALITY LEVEL) = 1.5. ALTHOUGH THIS CONTAMINATION LEVEL IS VERY LOW, WE CANNOT COMPLETELY EXCLUDE THAT A CUSTOMER MAY RECEIVE A CONTAMINATED PLATE. INVESTIGATION CONCLUSION: BASED ON THE EVALUATION OF THE PROVIDED REPORT AND ON THE BASIS OF THE PROVIDED PICTURE, THE COMPLAINT WAS CONFIRMED. A CORRECTIVE AND PREVENTIVE ACTION WILL NOT BE IMPLEMENTED AS A TREND COULD NOT BE IDENTIFIED. WE WOULD SUGGEST TO SET ASIDE, AND NOT USE, ANY PREPARED PLATED MEDIA THAT DOES NOT MEET THE APPEARANCE SPECIFICATION AS IT IS DESCRIBED ON THE BD CERTIFICATE OF ANALYSIS. THIS IS CONSISTENT WITH INDUSTRY RECOMMENDATIONS FOR INSPECTION OF CULTURE MEDIA PRIOR TO USE (E.G. ¿GOOD PRACTICES FOR PHARMACEUTICAL MICROBIOLOGY LABORATORIES¿, WHO TECHNICAL REPORT SERIES, NO. 961, 2011, ANNEX 2; CHAPTER <1117> ¿MICROBIOLOGY BEST LABORATORY PRACTICES¿ THE UNITED STATES PHARMACOPEIA; AND THE DIFCO & BBL MANUAL). H3 OTHER TEXT : SEE H.10.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US BUT IS CONSIDERED TO BE SUBSTANTIALLY SIMILAR TO THE LEGALLY U.S. MARKETED DEVICE BD BBL¿ (B)(6) AGAR WITH 5% SHEEP BLOOD CATALOG NUMBER 221263 WHICH IS A PREAMENDMENT DEVICE.
IT WAS REPORTED THAT WHILE USING PLATE COLUMBIA AG 5PRCT SB 90MM CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "7 OUT OF 140 PLATER ARE CONTAMINATED."
IT WAS REPORTED THAT WHILE USING PLATE (B)(6) AG 5PRCT SB 90MM CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "7 OUT OF 140 PLATER ARE CONTAMINATED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 976766 | PLATE COLUMBIA AG 5PRCT SB 90MM | CULTURE MEDIA, ENRICHED | KZI | BECTON DICKINSON GMBH | 1082404 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |