FDA Adverse Event Injury Summary report: N

GAP SPACER

MDR report key: 12076229 · Received June 28, 2021

Report

Report Number
1020279-2021-05462
Event Type
Injury
Date Received
June 28, 2021
Date of Event
June 4, 2021
Report Date
August 3, 2021
Manufacturer
SMITH & NEPHEW, INC.
Product Code
HWT
UDI-DI
00885556578445
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

H3, H6: THE DEVICE, USED IN TREATMENT, WAS NOT RETURNED FOR EVALUATION AND THE REPORTED EVENT COULD NOT BE CONFIRMED. AT THIS TIME, WE HAVE NO REASON TO SUSPECT THAT THE PRODUCT FAILED TO MEET ANY PRODUCT SPECIFICATIONS AT THE TIME OF MANUFACTURE. A COMPLAINT HISTORY REVIEW FOUND RELATED FAILURES; THIS FAILURE MODE WILL BE MONITORED FOR FUTURE COMPLAINTS AND ASSESSED FOR ANY NECESSARY CORRECTIVE ACTIONS. A REVIEW OF THE RISK MANAGEMENT FILE REVEALED THIS FAILURE MODE WAS PREVIOUSLY IDENTIFIED. THIS DEVICE IS A REUSABLE INSTRUMENT THAT CAN BE EXPOSED TO NUMEROUS SURGERIES. DAMAGE FROM PROLONGED USE, MISUSE OR ROUGH HANDLING ARE LIKELY PROBABLE CAUSES OF THE REPORTED EVENT. WE RECOMMEND THAT ALL REUSABLE INSTRUMENTS BE ROUTINELY INSPECTED FOR WEAR AND DAMAGE AND REPLACED AS NECESSARY. BASED ON THIS INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. WITHOUT THE RETURN OF THE ACTUAL PRODUCT INVOLVED OR PRODUCT INFORMATION, OUR INVESTIGATION COULD NOT PROCEED. SHOULD THE DEVICE OR ADDITIONAL INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. NO FURTHER INVESTIGATION IS WARRANTED FOR THIS COMPLAINT; HOWEVER, WE WILL CONTINUE TO MONITOR FOR FUTURE COMPLAINTS AND INVESTIGATE AS NECESSARY. WE CONSIDER THIS INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING A TKA, THE STICK OF A GAP SPACER SNAPPED WHEN USED TO ASSESS GAPS INSIDE THE PATIENT. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING THE SAME DEVICE. PATIENT WAS NOT HARMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
973574 GAP SPACER TEMPLATE HWT SMITH & NEPHEW, INC. 71935180 UNK 00885556578445

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R