FDA Adverse Event Malfunction Summary report: N

CHLORAPREP ONE STE HI LITE ORANGE

MDR report key: 12075648 · Received June 28, 2021

Report

Report Number
3004932373-2021-00304
Event Type
Malfunction
Date Received
June 28, 2021
Date of Event
April 23, 2021
Report Date
August 5, 2021
Manufacturer
CAREFUSION 213, LLC 0113
Product Code
KXG
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE FACILITY DID PROVIDE PHOTOS/SAMPLES TO AID IN OUR QUALITY ENGINEER¿S INVESTIGATION. WITH THE SAMPLE, PROVIDED, BD WAS ABLE TO CONFIRM THE FAILURE MODE AS THE END CAP WAS DETACHED FROM THE APPLICATOR BODY. A DEVICE HISTORY RECORD WAS REVIEWED, AND NO NON-CONFORMANCES WAS NOTED DURING THE MANUFACTURING OF THIS LOT. THE ROOT CAUSE IS ATTRIBUTED TO THE EQUIPMENT STATION FOR THE END CAP PLACEMENT UNTO THE APPLICATOR BODY. CORRECTIVE ACTIONS WERE INITIATED WHICH LED TO BD CONDUCTING A VOLUNTARY RECALL ON CERTAIN LOTS OF THE CHLORAPREP HI- LITE ORANGE 26 ML APPLICATOR. BD HAS CONFIRMED THAT SOME OF THE PRODUCT, WHICH INCLUDED THIS LOT, HAD AN APPLICATOR END CAP THAT WAS IMPROPERLY SECURED DURING THE MANUFACTURING PROCESS WHICH RESULTED IN BROKEN GLASS DROPPING OUT OF THE APPLICATOR. FOLLOW UP EMDR PR 3061775 H3 OTHER TEXT : PLEASE SEE NARRATIVE BELOW.

Description of Event or Problem · 0

MATERIAL NO.: 930815 . BATCH NO.: 0328213. IT WAS REPORTED BY THE USER FACILITY THAT THE END OF THE PREP STICK CAME OFF AND THE GLASS AMPULE BROKE ON THE FLOOR. PER MEDWATCH: DESCRIBE THE EVENT OR PROBLEM: CHLORAPREP OBTAINED FROM SUPPLY CABINET WHILE REMOVING FROM PACKAGE THE END OF PREP STICK CAME OFF GLASS AMPULE FELL TO FLOOR AND BROKE. LOT: 0328213 EXP: 11/2023 26ML ORANGE APPLICATOR NOT SAVED LOT# ABOVE IS ON THE BECTON-DICKMSON. VOLUNTARY RECALL LIST ISSUED 4/20/2021.

Additional Manufacturer Narrative · 1

MEDWATCH # 4900240000-2021-8071. (B)(4). PR (B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

MATERIAL NO.: 930815. BATCH NO.: 0328213. IT WAS REPORTED BY THE USER FACILITY THAT THE END OF THE PREP STICK CAME OFF AND THE GLASS AMPULE BROKE ON THE FLOOR. PER MEDWATCH: DESCRIBE THE EVENT OR PROBLEM: CHLORAPREP OBTAINED FROM SUPPLY CABINET WHILE REMOVING FROM PACKAGE THE END OF PREP STICK CAME OFF GLASS AMPULE FELL TO FLOOR AND BROKE. LOT: 0328213 EXP: 11/2023 26ML. ORANGE APPLICATOR NOT SAVED LOT# ABOVE IS ON THE BECTON-DICKMSON VOLUNTARY RECALL LIST ISSUED 4/20/2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
970969 CHLORAPREP ONE STE HI LITE ORANGE 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL KXG CAREFUSION 213, LLC 0113 0328213

Patients

Seq Age Sex Outcome Treatment
1 Other